La Crosse Technology INFRASCAN IR 101 User Manual
Page 5

5
5. Back light
The LCD back light is constantly switched ON while the operating button is 
pressed. When the operating button is released, the backlight will be ON for 
approximately 4 seconds before automatically switching off 8 seconds after the 
last measurement. 
 
6. Measurement inaccuracies 
The following points should be observed in order to avoid measurement errors:
a. It must be ensured that the object to be measured completely fills the
Infrared Thermometer monitoring area. Errors can occur when 
measuring small objects, as the unit also monitors the temperature of 
the surroundings as well as the object to be measured. 
b. The principle of measurement requires the ambient temperature to be
determined. This takes place at one-minute intervals or each time the 
unit is switched on. Changes in the ambient temperature, which can 
occur when the user holds the unit in their hand, are not monitored 
within this time period and therefore neither of them is taken into 
account when determining the indicated temperature. A gradual 
warming of the unit (the lens) between ambient temperature 
measurements can, for example, lead to the indicated temperature 
slowly falling. 
c. If the emission factor for the object does not correspond to the preset
value of 0.95, the temperature will not be calculated correctly. Most 
surfaces have an emission factor in the range of 0.8 to 0.98. In general, 
the darker and more matte the surface is, the larger their emissions 
factor. If the factor is greater the 0.95, the temperature indicated will 
be greater than the actual temperature of the object being measured. If 
the factor is less then 0.9, the temperature indicated will be less than 
the surface temperature. Polished metals have a very low emission 
factor and are therefore not suitable for measuring with an infrared 
thermometer. On the other hand, affixing adhesive tape to or painting 
the surface will increase the emission factor and reduce the inaccuracy 
of the measurement. 
d. It is not possible to take measurements through transparent materials
(glass, plexiglas, etc.).
e. It is not possible to measure air temperatures. 
f. 
Measurement errors can occur due to air contaminated with dust, 
steam, smoke, etc. 