beautypg.com

Chroma metercs-100a, Cs-100a, Chroma meter – Konica Minolta ILLUMINANCE METER LUMINANCE METER CHROMA METER T-10A /T-10MA/ T-10WSA/T-10WLA User Manual

Page 9

background image

16

17

A compact, lightweight, battery-powered

instrument with a 1° measurement angle

for high-accuracy non-contact measurements

of luminance and chromaticity

Easy-To-Read Display

Main Features

Compact and lightweight
Measurements of subjects at a distance

SLR (single-lens-reflex) viewing system and flare-free optical

system provide accurate measurements of subjects at a

distance with virtually no influence from light outside the

measurement area.

Measurements of small subjects

1° measurement angle allows measurements of subjects as

small as ø14.4 mm (at a subject distance of 1014 mm); by

using optional Close-Up Lenses, subjects as small as ø1.3 mm

can be measured.

Luminance units of cd/m

2

or fL can be

selected

Main Applications

Light-Source Measurements

• Luminance and chromaticity of small light sources such as

LEDs, miniature neon lamps, etc.

• Luminance and chromaticity of general light sources such

as tungsten lamps, fluorescent lamps, etc.

• Luminance and chromaticity of traffic signals, airport

guidance lights, emergency exit signs, etc.

Reflective-Subject Measurements

• Color measurements of subjects which cannot be measured

by contact methods, such as distant building walls, just-

painted surfaces.

Display Measurements

• Luminance and chromaticity of projectors.

Field of view (9°)

Measured luminance

(cd/m

2

or fL)

Measuring area (1°)

Measured value

Measurement mode

Viewfinder Image

External display

Specification

Model

Chroma Meter CS-100A

Type

SLR spot colorimeter for measuring light-source and surface luminance and chromaticity

Measuring angle

Optical system

85 mm f/2.8 lens; SLR viewing system; flare factor less than 1.5%

Angle of view

9° with 1° measurement area indication

Focusing distance

1014 mm (40 in.) to infinity

Receptors

3 silicon photocells filtered to detect primary stimulus values for red, green and blue light

Spectral response

Closely matches CIE 1931 Standard Observer curves (x¯

2

λ,y¯λ, and z¯λ)

Response time

Switchable; FAST: Sampling time: 0.1s, Time to display: 0.8 to 1.0s

SLOW: Sampling time: 0.4s, Time to display: 1.4 to 1.6s

Luminance units

cd/m

2

or fL (switchable)

Measuring range*

1

FAST: 0.01 to 299,000 cd/m

2

(0.01 to 87,530 fL)

SLOW: 0.01 to 49,900 cd/m

2

(0.01 to 14,500 fL)

Accuracy*

2

Luminance (Y): ±2% of reading ±1 digit

Chromaticity (x,y): FAST: 48.1 cd/m

2

or above: ±0.004; below 48.1 cd/m

2

: below measurement range

SLOW: 12.0 cd/m

2

or above: ±0.004; below 12.0 cd/m

2

: below measurement range

Repeatability*

1

Luminance (Y): ±0.2% of reading ±1 digit

Chromaticity (x,y) :

FAST: 100 cd/m

2

or above: ±0.001; 48.1 to 99.9 cd/m

2

: ±0.002; below 48.1 cd/m

2

: below measurement range

SLOW: 25.0 cd/m

2

or above: ±0.001; 12.0 to 24.9 cd/m

2

: ±0.002; below 12.0 cd/m

2

: below measurement range

Target value

1; set by measurement or numerical input

Measurement modes

Absolute color: Yxy; color difference: (Yxy)

Display

External: LCD; 3 values (Y, x, and y) of 3 digits each with additional indications

Viewfinder: 3-digit LCD (showing luminance value Y) with LED backlight

Data communication

RS-232C; baud rate: 4800 bps

External control

Measurement process can be started by external device connected to data output terminal

Power source

One 9 V battery; power can also be supplied via data output terminal

Operating temperature/humidity range

0 to 40°C, relative humidity 85% or less (at 35°C) with no condensation

Storage temperature /humidity range

-20 to 55°C , relative humidity 85% or less (at 35°C) with no condensation

Dimensions

79x208x154 mm (3-1/8x8-3/16x6-1/16 in.)

Weight

890 g (2 lb.) without battery

Standard accessories

Lens cap; Eyepiece cap; Protective filter, ND eyepiece filter; 9V battery; Chromaticity chart; Case

*1 Standard Illuminant A

*2 Standard Illuminant A measured at ambient temperature of 20 to 30ºC

Optional Accessories

Close-Up Lenses

Angle Finder V

N

Angle Finder V

N

allows the measuring

area and measurement display inside

the viewfinder to be seen at an angle of

90° to the normal viewfinder optical axis.

Angle Finder V

N

can also be focused and

the magnification can be set to 1x or 2x.

Data Management Software

CS-S10w Professional (Optional accessory)

Color space

: Lv x y, Lv u’ v’, LvTuv , XYZ, dominant wavelength

Mode selection : Normal mode, Object color mode, Contrast mode

RGB mode, RGB & contrast mode

Instrument control : Average measurement, Interval measurement

Data management : Reading and saving files, Data management with folders

Creating, saving and loading templates

(customizable design/layouts for various graphs)

Various graph displays

Data evaluation : Statistics display for each folder

Box tolerance setting, Multiple-point

measurement, uniformity display, contrast

display and polygon tolerance setting for display

evaluation

Other

: Creating reports in customizable screen layouts

System requirements

OS

Windows

®

XP Professional 32-bit SP3, 64-bit SP2

Windows

®

Vista Business 32-bit, 64-bit,

Windows

®

7 Professional 32-bit, 64-bit

CPU

Pentium

®

III 600 MHz equivalent or higher

Memory 128 MB min. (256 MB or more recommended)

Hard disk 60 MB or more space required for installation

Display

1024 X 768, 256 colors or more

Other

CD-ROM drive, USB port

Close-Up Lenses

Minimum

measuring area

No.153

ø8.0 mm

No.135

ø5.2 mm

No.122

ø3.2 mm

No.110

ø1.3 mm

(Theoretical values)

System diagram

(Optional Accessories)

Close-Up Lenses

No. 153

No. 135

No. 122

No. 110

Angle Finder V

N

White Calibration

Plate (for 45-0)

PC

Connecting Cable

LS-A12

Data Management Software

CS-S10w

Professional

Dimensions

(Units: mm)

38

19

208

176

85.8

151

58 49.6

154.3

57

40

28.5

16

60.3

44

78.7

Tripod socket

Standard plane

for distance

measurements

Chroma Meter

CS-100A

Chroma Meter CS-100A

Chroma Meter CS-100A