Specifications, Mdf-u7386s, Dimensions – Sanyo MDF-793C User Manual
Page 4: Performance data options, Features
w w w . s a n y o b i o m e d i c a l . c o m
Specifications
SANYO Commercial Solutions
A Division of SANYO North America
1062 Thorndale Avenue, Bensenville, IL 60106 USA
Toll Free USA 800-858-8442 • Fax 630-238-0074
www.sanyobiomedical.com
SANYO Canada, Inc.
1- 300 Applewood Crescent, Concord, Ontario L4K 5C7
905-760-4025 • Fax 905-760-9945
DMVER606A
p
r
e
s
e
r
v
a
t
i
o
n
Specifications
Model
Temperature range
Exterior dimensions
(W x D x H)
Interior dimensions
(W x D x H)
Effective capacity
Exterior Cabinet
Interior Cabinet
Inner Lid
Insulation
Compressor High stage side
Low stage side
Evaporator High stage side
Low stage side
Condenser High stage side
Low stage side
Temperature Control
Temperature Display
Sensor
Alarm System
Net Weight (Approx.)
Power Requirement
MDF-593C
-20°C to -86°C
2,010 x 770 x 1,070
(79.1 x 30.3 x 42.1)
1,280 x 500 x 762
(50.3 x 19.7 x 30)
487 liters (17.1 cu.ft.)
3
Hermetic type 750W
Hermetic type 750W
305kg (672 lbs)
*Appearance and specifications are subject to change without notice.
Painted Steel
Stainless Steel
Cascade condenser
Tube on sheet (shared with interior)
Fine and tube type
Shell and tube type
Microprocessor: Keypad input
Temp. input range: -20°C to -95°C (1C increment)
Set value memory: non-volatile memory
Digital Display
Platinum resistance (Pt. 100)
Selectable high temp. alarm (+10°C & +15°C from set point), Power Failure alarm, Filter
check lamp, Remote alarm contact
MDF-793C
-20°C to -86°C
2,570 x 770 x 1,070
(101.2 x 30.3 x 42.1)
1,840 x 500 x 762
(72.4 x 19.7 x 30)
701 liters (24.7 cu.ft.)
4
Hermetic type 1,100W
Hermetic type 1,100W
375kg (827 lbs)
Rigid polyurethane foamed-in place
220V / 60hz Built-in voltage booster
Specifications
Model
Temperature range
Maximum cooling
performance
Exterior dimensions
(W x D x H)
Interior dimensions
(W x D x H)
Net weight
Effective capacity
Shelf
Access port
Compressor
Refrigerant
Alarm
Remote alarm contact
Accessories
Power Requirement
MDF-U7386SC
–50°C to –86°C (1°C increment)
–86°C
(Ambient temp. 30°C)
44.5” x 34.4” x 78.3”
(1130 x 875 x 1990mm)
34.2” x 23.6” x 50.4”
(870 x 600 x 1280mm)
783 lbs. (Approx. 355kg)
23.5 cu.ft. (668 L)
Stainless steel, Adjustable,
3 shelves, W848 x D533mm,
Max. load: 50kg (110 lbs.)/shelf
17mm diameter, 3 locations
(back, bottom left/right corner)
Hermetic type, Output: 1100 W
(high stage), 1100 W (low stage)
HFC refrigerants
High/low temperature,
Power failure, Filter check,
Battery check
Allowable contact capacity:
DC 30V, 2A
1 set of keys, 1 scraper
220V / 60 Hz
Built-in voltage booster
MDF-U5386SC
–50°C to –86°C (1°C increment)
–86°C
(Ambient temp. 30°C)
35.0” x 34.4” x 78.3”
(890 x 875 x 1990mm)
24.8” x 23.6” x 50.4”
(630 x 600 x 1280mm)
672 lbs. (Approx. 305kg)
17.1 cu.ft. (483 L)
Stainless steel, Adjustable,
3 shelves, W608 x D533”, Max.
load: 50kg (110 lbs.)/shelf
.07”(17mm) diameter, 3 locations
(back, bottom left/right corner)
Hermetic type, Output: 1100 W
(high stage side), 1100 W (low
stage side)
HFC refrigerants
High/low temperature,
Power failure, Filter check,
Battery check
Allowable contact capacity:
DC 30V, 2A
1 set of keys, 1 scraper
220V / 60 Hz/ 1 phase
Built-in voltage booster
Nema 6-15 Recepticle
*Appearance and specifications are subject to change without notice.
MDF-U7386S
Dimensions
[Unit : mm (inch)]
Specifications
Model
Temperature range
Maximum cooling
performance
Exterior dimensions
(W x D x H)
Interior dimensions
(W x D x H)
Net weight
Effective capacity
Shelf
Access port
Compressor
Refrigerant
Alarm
Remote alarm contact
Accessories
Power Requirement
MDF-U7386SC
–50°C to –86°C (1°C increment)
–86°C
(Ambient temp. 30°C)
44.5” x 34.4” x 78.3”
(1130 x 875 x 1990mm)
34.2” x 23.6” x 50.4”
(870 x 600 x 1280mm)
783 lbs. (Approx. 355kg)
23.5 cu.ft. (668 L)
Stainless steel, Adjustable,
3 shelves, W848 x D533mm,
Max. load: 50kg (110 lbs.)/shelf
17mm diameter, 3 locations
(back, bottom left/right corner)
Hermetic type, Output: 1100 W
(high stage), 1100 W (low stage)
HFC refrigerants
High/low temperature,
Power failure, Filter check,
Battery check
Allowable contact capacity:
DC 30V, 2A
1 set of keys, 1 scraper
220V / 60 Hz
Built-in voltage booster
*Appearance and specifications are subject to change without notice.
ISO 9001 & 14001 certified
Distributed by:
Performance Data
Options
Circular recorder
MTR-C954
Strip Chart Recorder
MTR-85H
Recorder mounting
bracket
MDF-S3085
(for MTR-85H)
LCO
2
backup system
CVK-UB2 (I)
LN
2
backup system
CVK-UBN2
40
20
0
–20
–40
–60
–80
–100
0
2
4
6
8
10
5
0
12
35
25
20
14
15
10
Te
m
p
er
at
u
re
(°
C
)
40
20
0
–20
–40
–60
–80
–100
0
2
4
6
8
10
12
14
Te
m
p
er
at
u
re
(°
C
)
Time (hour)
Time (hour)
Pull-up characteristics
during power failure
30
12
8
6
4
2
10
0
T
im
e
(h
o
u
r)
Ambient temperature (°C)
CO
2
backup system operation
characteristics
Central air temp. in cabinet
Ambient temp. 30°C, No-load
Ambient temp. and maintenance
of –70°C by injection of LCO
2
(30kg)
Inventory rack
SUR-524-I
Rack Capacity: 24
SANYO Electric Co., Ltd.
SANYO Sales & Marketing Corporation
5-15, Hiyoshi-cho 2-chome, Moriguchi City, Osaka, 570-8634, Japan
Telephone: +81-6-6992-0322 Fax: +81-6-6992-9291
URL http://www.sanyo.co.jp/cmg/biomedical
©2006 SANYO Printed in Japan 2006.3. MA. SHR101
18
(0.7)
48
(1.9)
1130 (44.5)
870 (34.3)
1196 (47.1)
1540 (60.6)
450 (17.7)
1990 (78.3)
605.5 (23.8)
105
(4.1)
1280 (50.4)
130 (5.1)
130 (5.1)
140
(5.5)
795 (31.3)
867 (34.1)
11
(0.4)
1625 (64)
Features
EVOLUTIONARY DESIGN
• The newly developed refrigeration system
and the freezer structure offer a quiet
experimental environment.
• Achieves 10% higher energy conservation
than previous model.
(For model rated 220V, 50Hz)
• The newly developed back cover combined
with new aerodynamically designed and
placed components in the refrigeration
compartment provide superior air flow,
making it possible
to drastically
reduce the stress
to the freezer and
contributing to
excellent durability.
• Two independent and insulated inner doors
prevent cold air leakage.
ACCESSIBILITY
• A new beak style inner
door latch tightly closes
the inner door against
the freezer frame. It
also helps make
opening and closing the door smoother.
• The condenser filter is situated at the
bottom right side of the front panel to make
filter removal and cleaning easier.
Back
cover
Compressors
Front
Rear
Air
SAFETY
• Alarm lamp and buzzer offer secure warning
of power failure or abnormal temperature
increase.
• High and low temperature warning provides
an audible and visual alarm when the
temperature deviates more than 5°C to
20°C (adjustable) from the set point.
• Alarm ring back function ensures that buzzer
will resume operation, should alarm
conditions continue after it is silenced.
• Microprocessor-controlled filter-clog check
function protects the refrigeration circuit.
• New rugged, one-handed
outer door latch has a hole to
allow a padlock to securely
protect valuable samples.
• Control panel with
digital display for easy
operation.
Pull-down characteristics
*Cooling performance is indicated by the temperature reached at
the center of the freezer (at ambient temperature of 30C with no
load). In order to use the freezer at a stable temperature for a long
time, it is recommended that the temperature be set to at least 5C
higher than the indicated lowest temperature.
In addition, depending on the usage conditions, it may not be