beautypg.com

Siemens SITRANS F M MAGFLO MAG 5100 W User Manual

Page 6

background image

6

SITRANS F M MAGFLO

Electromagnetic flowmeter type MAG 5100 W

mm

inch

Nm

f/lbs

Nm

f/lbs

Nm

f/lbs

Nm

f/lbs

Nm

f/lbs

25

1"

N/A

N/A

N/A

N/A

10

7

7

5

N/A

N/A

40

1½"

N/A

N/A

N/A

N/A

16

12

9

7

N/A

N/A

50

2"

N/A

N/A

25

18

N/A

N/A

25

18

N/A

N/A

65

2½"

N/A

N/A

25

18

N/A

N/A

25

18

N/A

N/A

80

3"

N/A

N/A

25

18

N/A

N/A

34

25

N/A

N/A

100

4"

N/A

N/A

25

18

N/A

N/A

26

19

N/A

N/A

125

5"

N/A

N/A

29

21

N/A

N/A

42

31

N/A

N/A

150

6"

N/A

N/A

50

37

N/A

N/A

57

42

N/A

N/A

200

8"

50

37

50

37

N/A

N/A

88

65

N/A

N/A

250

10"

50

37

82

61

N/A

N/A

99

73

N/A

N/A

300

12"

57

42

111

82

N/A

N/A

132

97

N/A

N/A

350

14"

60

44

120

89

N/A

N/A

225

166

N/A

N/A

400

16"

88

65

170

125

N/A

N/A

210

155

N/A

N/A

450

18"

92

68

170

125

N/A

N/A

220

162

N/A

N/A

500

20"

103

76

230

170

N/A

N/A

200

148

N/A

N/A

600

24"

161

119

350

258

N/A

N/A

280

207

N/A

N/A

700

28"

200

148

304

224

N/A

N/A

N/A

N/A

200

148

750

30"

N/A

N/A

N/A

N/A

N/A

N/A

N/A

N/A

240

177

800

32"

274

202

386

285

N/A

N/A

N/A

N/A

260

192

900

36"

288

213

408

301

N/A

N/A

N/A

N/A

240

177

1000

40"

382

282

546

403

N/A

N/A

N/A

N/A

280

207

42"

N/A

N/A

N/A

N/A

N/A

N/A

N/A

N/A

280

207

1100

44"

N/A

N/A

N/A

N/A

N/A

N/A

N/A

N/A

290

214

1200

48"

395

292

731

539

N/A

N/A

N/A

N/A

310

229

AWWA

Nominal size

PN 10

PN 16

PN 40

Class 150

Maximum allowable
torques

All values are theoretical and are calculated making the following assumptions:

1) All bolts are new and material selection is according to EN 1515-1 table 2

2) Gasket material not exceeding 75 shore A durometer is used between the flowmeter and

mating flanges

3) All bolts are galvanized and adequately lubricated

4) The values are calculated for use with carbon steel flanges

5) Flowmeter and mating flanges are correctly aligned

Torque calculations

Standard bolts must be well lubricated and tight-
ened evenly around the gasket. Leakage/dam-
age to the flowmeter or piping may arise if bolts
are overtightened.