Mdf-u3386s, Dimensions, Accessibility – Sanyo MDF-U3386S User Manual
Page 2: Performance data safety, Backup system operation characteristics
MDF-U3386S
Dimensions
[Unit : mm (inch)]
Specifications
Features
Options
EVOLUTIONARY DESIGN
• The newly developed refrigeration
system and the freezer structure offer a
quiet experimental environment.
• Achieves 22% higher energy conser-
vation than previous model.
(230V, 50Hz)
• The newly developed back cover
combined with new aerodynamically
designed and placed components in the
refrigeration compartment provide
superior air flow, making it possible to
drastically reduce
the stress to the
freezer and
contributing to
excellent
durability.
• Two independent and insulated inner
doors prevent cold air leakage.
Back
cover
Compressors
Front
Rear
Air
Model
Temperature range
Maximum cooling
performance
Exterior dimensions
(W x D x H)
Interior dimensions
(W x D x H)
Net weight
Effective capacity
Shelf
Access port
Compressor
Refrigerant
Alarm
Remote alarm contact
Accessories
MDF-U3386S
–50˚C to –86˚C (1˚C increment)
–86˚C
(Ambient temp. 30˚C)
750 x 875 x 1850mm
(29.5” x 34.4” x 72.8”)
490 x 600 x 1140mm
(19.3” x 23.6” x 44.9”)
Approx. 255 kg (562 lbs.)
333 L (11.8 cu.ft.)
Stainless steel, Adjustable,
3 shelves, Max. load: 50kg
(110 lbs.)/shelf
17mm diameter, 3 locations
(back, bottom left/right corner)
Hermetic type, Output: 450W
(high stage side), 750W (low
stage side)
HFC refrigerant
High/low temperature, Power
failure, Filter check, Battery
check, Part replacement
notification
Allowable contact capacity:
DC 30V, 2A
1 set of keys, 1 scraper
•Appearance and specifications are subject to change without notice.
ACCESSIBILITY
• A new beak style
inner door latch tightly
closes the inner door
against the freezer
frame. It also helps make opening and
closing the door smoother.
• The condenser filter is situated at the
bottom right side of the front panel to
make filter removal and cleaning easier.
Performance Data
SAFETY
• Alarm lamp and buzzer offer secure
warning of power failure or abnormal
temperature increase.
• High and low temperature warning
provides an audible and visual alarm
when the temperature deviates more
than
±5˚C to ±20˚C (adjustable) from the
set point.
• Alarm ring back function ensures that
buzzer will resume operation, should
alarm conditions continue after it is
silenced.
• Microprocessor-controlled filter-clog
check function protects the refrigeration
circuit.
• New rugged, one-handed
outer door latch has a hole to
allow a padlock to securely
protect valuable samples.
• Control panel with digital
display for easy operation.
*Cooling performance is indicated by the temperature reached at
the center of the freezer (at ambient temperature of 30
°C with no
load). In order to use the freezer at a stable temperature for a long
time, it is recommended that the temperature be set to at least
5
°C higher than the indicated lowest temperature.
In addition, depending on the usage conditions, it may not be
possible to reach the indicated lowest temperature.
Inventory rack
IR-216U
Inventory rack
IR-220U
Rack Capacity
IR-216U
IR-220U
IR-312UN
For 2-inch box
For 3-inch box
6
6
12
Circular recorder
MTR-G85
LCO
2
backup system
CVK-UB2
SANYO DAQ (data acquisition) System
MTR-5000/
Ethernet (LAN) Interface
MTR-L03
or RS232C/RS485 Interface
MTR-480
Strip chart recorder
MTR-85H
Recorder mounting
bracket
MDF-S3085
(for MTR-85H)
40
20
0
–20
–40
–60
–80
–100
0
2
4
6
8
10
10
5
12
35
30
25
14
20
15
T
e
mperature (˚C)
40
20
0
–20
–40
–60
–80
–100
0
2
4
6
8
10
12
14
T
e
mperature (˚C)
Time (hour)
Time (hour)
Pull-up characteristics
during power failure
Pull-down characteristics
20
15
10
5
T
ime (hour)
Ambient temperature (˚C)
CO
2
backup system operation
characteristics
Central air temp. in cabinet
Ambient temp. 30˚C, No-load
Ambient temp. and maintenance
of –70˚C by injection of LCO
2
(30kg)
18
(0.7)
750 (29.5)
1850 (72.8)
1400 (55.1)
450 (17.7)
490 (19.3)
48
(1.9)
816 (23.1)
1140 (44.9)
600 (23.6)
105
(4.1)
875 (34.4)
11
(0.4)
795 (31.3)
1485 (58.5)
130 (5.1)
130 (5.1)
140
(5.5)
38
(1.5)
34
(1.3)
218 (8.6)
Caution:
SANYO guarantees the product under certain warranty conditions.
SANYO in no way shall be responsible for any loss of content or
damage to content.
Distributed by:
©2009 SANYO Printed in Japan 2009.9 MA
SHR129
SANYO Electric Co., Ltd.
Biomedical Division
http://biomedical.sanyo.com/
SANYO Electric Co., Ltd., Biomedical Division, Gunma is certified for:
Quality management system: ISO9001 / Medical devices quality management system: ISO13485 / Environmental management system: ISO14001