SECTION 3—CLEANING AND DISINFECTING
21
Ꮨ DANGER
DO NOT attempt to separate the outer case of this unit.
Removing the case creates a serious electrical shock
hazard. All disassembly and maintenance of this unit
MUST be done by a qualified technician.
Cleaning the Nebulizer
ƽ WARNING
To prevent possible risk of infection due to contamination,
follow these instructions:
• Clean the nebulizer before initial operations.
• Clean the nebulizer after each aerosol treatment.
• ALWAYS use fresh cleaning solution.
• DO NOT towel dry the nebulizer components after
cleaning.
NOTE: Your physician and/or an Invacare dealer may specify
certain cleaning procedures. If so, follow their recommendations. If
not, clean according to the following instructions:
1. Remove the nebulizer and mouthpiece from the tubing.