beautypg.com

Section 7- troubleshooting guide, Symptom, Probable cause – Invacare IRC5P User Manual

Page 28: Solution, Section 7—troubleshooting guide, Symptom: probable cause: solution

background image

SECTION 7—TROUBLESHOOTING GUIDE

Perfecto

2

Series

28

Part No 1143482

SECTION 7—TROUBLESHOOTING
GUIDE

SYMPTOM:

PROBABLE CAUSE:

SOLUTION:

Alarm:

Main Power Loss:

Short beeps, long
pause

1. Power cord not
plugged in.

1. Insert plug into outlet.

Concentrator not
operating,
power switch On.

2. No power at outlet.

2. Inspect house circuit
breakers or fuses. If problem
recurs, use a different outlet.

Beep......Beep......

3. Tripped circuit
breaker.

3. Call Service Provider.

Alarm:
Continuous

Concentrator not
operating, Power
Switch On.

Beeeeeeeep....

System Failure:
1. Unit overheating due
to blocked air intake.

2. Insufficient power at
outlet.

3. Internal repairs
required.

1a. Remove and clean cabinet
filters.

1b. Move oxygen
concentrator at least three
inches away from walls,
draperies or furniture.

2. DO NOT use extension
cords. Move to another
electrical outlet or circuit.

3. Call Service Provider.

This manual is related to the following products: