Cleaning the mouthpiece, Changing the nebulizer, Cleaning the compressor – Invacare IRC 1730 User Manual
Page 29: Ꮨ danger

SECTION 4—CLEANING AND DISINFECTING
29
Cleaning the Mouthpiece
1. Boil the mouthpiece in water for a maximum of five
minutes or autoclave at a maximum of 121° C for ten
minutes.
2. Allow the mouthpiece to air dry.
3. Once dry, the mouthpiece may be stored inside the
nebulizer top.
Changing the Nebulizer
The nebulizer is designed for single patient use.
The nebulizer MUST be replaced after a long period of
inactivity, or when it is obstructed by dry medication, dust,
etc. The reusable nebulizer should last from six months to one
year with general use. The disposable nebulizer should last 30
days.
Cleaning the Compressor
Ꮨ DANGER
DO NOT remove outer case of this unit. Removing the
case creates a serious electrical shock hazard. All
disassembly and maintenance of this unit MUST be done
by a qualified technician.