SPECIFICATIONS
Electrical Requirements:
90-240 VAC, 50-60 Hz, or
12 -14 VDC
Ambient Temperature
Range:
10
o
C - 40
o
C (50
o
F - 104
o
F)
NOTE: Operating at extreme temperatures may increase treatment times.
Power Wattage:
24 Watts, @ 11 ± 1 PSI
Thermal Protector:
Internal Solid State - Auto Resettable
NOTE: The Freeway Freedom unit and Battery Pack each have an internal
resettable thermal fuse in case of overheating (operating longer than 20
minutes) or using incorrect polarity DC jack.
Safety Systems:
Class II - Double Insulated
Type BF
Fuse:
T3 Amp Fuse
Intake Filter:
300 micron, polyethylene
Flowrate:
Flowrate (typical): 6 lpm with Sidestream
nebulizer-model # MS2400 (back pressure
approximately 9.5 +/- .5 PSI)
Width:
9.25-inches
Height:
6.5-inches
Depth:
4.5-inches (7.5-inches w/accessory bag)
*Weight:
3.4 lbs (5.2 lbs. w/all accessories)
Battery Pack Dimensions:
3.75 x 1.4 x 2.4-inches
IMPORTANT DC Inlet
Polarity:
External Positive + - Internal Negative
* Weights given include the 1.1 lb. removable battery.
Stands for Direct Current