Troubleshooting, Display codes – Invacare Jasmine JLS5 User Manual
Page 7

Part No 1152408
7
11. Remove the weight. Press the LOCK/UNLOCK key.
NOTE: The scale is now in test mode.
12. Press the ZERO key. The display reads 0.00 lb.
NOTE: The scale will now be in normal operation.
FIGURE 5 Calibrate the Jasmine Scale
Troubleshooting
Display Codes
CALIBRATION REQUIRED ‐ Indicates improper stored calibration data, calibration is necessary.
OVER CAPACITY ‐ Indicates a weight exceeding the capacity has been loaded on the scale.
SYMPTOM
PROBABLE CAUSE
SOLUTION
Unit does NOT work properly.
Battery failure.
Check battery. Replace if necessary.
Battery has been replaced and unit still does
NOT work properly.
Contact Invacare.
CAL button
MODEL JLS5
CAPACITY 500 lb, 227 kg
BACK VIEW OF SCALE DISPLAY
FRONT VIEW OF SCALE DISPLAY
Display Window
CAL
OCAP