12. Ensure that the status indicator on the replacement inflator cap is

green, the lanyard for the beaded manual inflation handle is present,
and the locking ring is locked.

13. To test for leakage, inflate the LP using the oral inflator tube – see

Inflating the LP. To avoid damage to the LP do not exceed 2 psi. After
two hours there should be no noticeable drop in pressure in the LP.

RePacking the lP

note: after each field-use inflation, the MD1250 lPs must be rearmed,
leak tested, and fully deflated before repacking.

1. Visually inspect the status indicator on

the hydrostatic inflator (Figure 13). Green
indicates that the inflator is armed and
ready for use. Red indicates that the inflator
has been fired and must be replaced – see
Re-arming the LP.

2. Visually inspect the expiry date on the

hydrostatic inflator (Figure 13). If the expiry
date has passed, replace the inflator – see
Re-arming the LP.

gas Removal from the inflation cell

WaRning: all gas must be evacuated from the inflation cell prior to
packing the lP. failure to follow this procedure will significantly degrade
the performance of the automatic inflation system.

1. A vacuum source capable of achieving at least 5 in-Hg and a vacuum

fitting suitable to connect the vacuum source to the oral inflator is
required to complete this procedure.

note: a 3 hP shop vaccuum will provide more than 5 in-hg of vaccuum.

2. Place the LP on a flat surface with the hydrostatic inflator facing upwards.
3. Attach the automatic/manual conversion cap (MA7219) to the

hydrostatic inflator.

4. Turn the LP upside down so that the hydrostatic inflator is facing

downwards.

5. Locate the oral inflation tube and turn the locking ring clockwise to

open the inflator valve and unlock the mouthpiece.

6. Attach a vacuum fitting to the oral inflator.
7. Turn on the vacuum source.
8. Pinch the black flange at the base of the oral inflation tube. This will

help prevent the bladder fabric from sealing off the vacuum.

9. Depress the mouthpiece to evacuate the entrapped gas from the

inflation cell.

Status

Indicator

Hydrostatic Water

Inlet Valve

Manual Inflation

Lanyard
Beaded Manual

Inflation Handle

Tabs

Black Locking Ring

Expiry Date

Figure 13

Tactical Inflatable Side Pouch LP

Owners Manual

7