3 specifications and environmental conditions 11, 1 specifications 11, 2 environmental conditions 11 – Fluke 9112A User Manual

Page 14: 3 warranty 11, Specifications and environmental conditions, Specifications, Environmental conditions, Warranty, 3specifications and environmental conditions, 1 specifications

background image

3

Specifications and Environmental
Conditions

3.1

Specifications

Operating Range

300°C to 1100°C

Stability

300°C

±0.05°C

500°C

±0.05°C

700°C

±0.1°C

1000°C

±0.1°C

1100°C

±0.1°C

Uniformity

300°C

±0.05°C

500°C

±0.08°C

700°C

±0.2°C

1000°C

±0.25°C

1100°C

±0.3°C

Stabilization Time

Typically 2 hours midrange, slower at the low temperature end (4
hours), faster at the high temperature end

Heater Power

3700 Watts High

Power Requirements

230 VAC (±10%), 50/60 Hz, 20 A

Outside Dimensions

18” H x 14.25”W x 26”D (457mm x 362mm x 660mm)

Weight

72.5 lbs

3.2

Environmental Conditions

Although the instrument has been designed for optimum durability and trou-
ble-free operation, it must be handled with care. The instrument should not be
operated in an excessively dusty or dirty environment. Maintenance and clean-
ing recommendations can be found in the Maintenance Section of this manual.

The instrument operates safely under the following conditions:

• temperature range: 5 - 40°C (41 - 104°F)
• ambient relative humidity: 15 - 50%
• pressure: 75kPa - 106kPa
• mains voltage within ± 10% of nominal
• vibrations in the calibration environment should be minimized
• altitude less than 2000 meters
• indoor use only

3.3

Warranty

Fluke Corporation, Hart Scientific Division (Hart) warrants this product to be
free from defects in material and workmanship under normal use and service

11

3 Specifications and Environmental Conditions

Specifications