Problem

Possible Causes and Solutions

The display shows any
of the following: err 1,
err 2, err 3, err
4, Err 5, Err 6, or
Err 7

Controller problem. The error messages signify the following problems with
the controller.

Err 1 - a RAM error
Err 2 - a NVRAM error
Err 3 - a Structure error
Err 4 - an ADC setup error
Err 5 - an ADC ready error
Err 6 – a defective control sensor
Err 7 – a heater error
Initialize the system by performing the Factory Reset Sequence describe
above.

Temperature cannot be
set above a certain point

Incorrect High Limit parameter. The High Limit parameter may be set below
400°C. Check this value as described in the 2200 User’s Guide.