Specifications – Fluke 430 Series User Manual
Page 5

Specifications
continued
Accuracy, resolution and range
Volt/Amps/Hertz
Measurement range
Resolution
Accuracy
Vrms (AC+DC)
1 to 1000 Vrms
0.1 Vrms
± 0.5 % of nominal voltage
Vpk
1 to 1400 Vpk
1 V
5 % of nominal voltage
CFV
1.0 to > 2.8
0.1
± 5 %
Arms (AC+DC)
0 to 20,000 Arms
1
0,001 to 10 Arms
1
± 1 % of reading ± 5 counts
3
with included clamps
0 to 400 Arms
0.1 and 1 Arms
Apk
using 1mV/A scaling
0 - 5500 Apk
10 A
± 5 %
CFA
using 1mV/A scaling
1 to 10
0.1
± 5 %
50 Hz nominal
42.50 to 57.50 Hz
0.01 Hz
± 0.1 % of nominal frequency
60 Hz nominal
51.00 to 69.00 Hz
0.01 Hz
± 0.1 % of nominal frequency
Dips and swells
Vrms
1
⁄
2
(AC+DC)
0.0 to100 % of nominal voltage
0.1Vrms
± 1 % of nominal voltage
Arms
1
⁄
2
(AC+DC)
0 to 20,000 Arms
1
0,001 Arms to 10 Arms
± 1 % of reading ± 5 counts
3
with included clamps
0 to 400 Arms
0.1 Arms and 1 Arms
Threshold levels
Programmable in percent of nominal voltage
Duration
hhh,mm,ss,mmm with half-cycle cycle time resolution
Harmonics
Harmonic selection (n)
DC, 1 to 50
Inter-Harmonic selection
Off, 1 to 49
Vrms
Relative (%f):
0.0 to 100.0 %
0.1 %
± 0.1 % ± n x 0.1 %
(± 0.4 % for %r)
Absolute:
0.0 to 1000 Vrms
0.1 Vrms
± 5 % of reading ± 2 counts
Arms
Relative (%f):
0.0 to 100.0 %
0.1 %
± 0.1 % ± n x 0.1 %
(± 0.4 % for %r)
Absolute:
0.0 to 4000 mV x selected
1 mVrms x selected clamp scaling
± 5 % of reading ± 5 counts
clamp scaling
Watts
Relative:
0.0 to 100.0 %
0.1 %
± n x 2 %
Absolute:
depends on selected clamp
± 5 % of reading ± n x 2 %
and voltage scaling
of reading, ± 10 counts
DC voltage
Relative:
0.0 to 100.0 %
0.1 %
± 1 %
Absolute:
0.0 to 1000 V
0.1 V
± 5 % of reading ± 10 counts
THD
0.0 to 100.0 %
0.1 %
± 2.5 %
Hz
0 to 3500 Hz
1 Hz
± 1Hz
Phase angle
-360 º to +360 º
1 º
± n x 1.5 º
Power and Energy
Watt
1.0 W to 20.00 MW
1
0.1 W to 1 kW
1
± 1.5 % of reading ± 10 counts
3
VA
1.0 VA to 20.00 MVA
1
0.1 VA to 1 kVA
1
± 1.5 % of reading ± 10 counts
3
VAR
1.0 VAR to 20.00 MVAR
1
0.1 VAR to 1 kVAR
1
± 1.5 % of reading ± 10 counts
3
kWh
00.00 kWhr to 200.0 GWhr
1
0.01 Whr to 100 Whr
1
± 1.5 % of reading ± 10 counts
3
kVA
00.00 kVAhr to 200.0 GVAhr
1
0.01 VAhr to 100 VAhr
1
± 1.5 % of reading ± 10 counts
3
kVAR
00.00 kVARhr to 200.0 GVARhr
1
0.01 VARhr to 100 VARhr
1
± 1.5 % of reading ± 10 counts
3
Power Factor
0 to 1
0.01
± 0.03
3
Cos
ϕ / DPF
0 to 1
0.01
± 0.03
3
Flicker
Pst (1min), Pst, Plt, PF5
0.00 to 20.00
0.01
Within ± 5 % of tabulated values
instantenous Flicker
according IEC61000-4-15
Dc %, Dmax % and Time d(t)
0.0 to ± 100.0 % for Dc % and
0.1 % for Dc % and Dmax % and
± 1% for Dc % and Dmax % and
exceeds limits. As described
Dmax % and 0.000 to 9.999s
10 ms for Time
20 ms for Time
per IEC 61000-3-3
for Time
Unbalance
Volts
0.0 to 5.0 %
0.1 %
± 0.5 %
Current
0.0 to 20 %
0.1 %
± 1 %
Transient capture
Volts
± 6000 Vpk
1 V
± 15 % of cursor reading
± 2.5 % of Vrms
Minimum detect duration
5 µs
Sampling rate
200 kS/s
Inrush mode
Arms (AC+DC)
0.000 to 20.00 kArms
1
0.001 to 10 Arms
1
± 1 % of meas ± 5 counts
3
Inrush Duration
mm:ss:mmm between 7.5 s to
10 ms
± 20 ms (Fnominal = 50 Hz)
30 m selectable
Fluke 430 Series Three-Phase Power Quality Analyzers
Fluke Corporation 5