F300S/F300SQ Ultraviolet Lamp System
NOTE: A dry contact (such as a relay contact or a PLC containing a relay
contact) with gold contact material should be used to drive the remote inputs.
The cable length to the dry contact should be no more than 6 meters. Shielded
cable should be used to minimize possible electrical interference.
In a multi-lamp system, connections are made to the Master power supply.
If there is a requirement for clearing faults remotely, a Remote Reset switch must
be used.
Output Ports (J105B)
There are five relay outputs available on the P300M/P300MQ/P300MT through
J105B. These outputs may be used to synchronize customer equipment with the
system. In a multi-lamp system, connections are made to the Master power supply.
Table 6. J106 Inputs (Remote Operation)
Signal
Pins
Active
Condition
Contact
Ratings
Application Comments
REMOTE
LAMP ON
4,5
Closed circuit
12mA @12 VDC
Set switch S1-2 of Master
to NO. Leave S1-1 and
S1-5 set to YES.
REMOTE
RESET
1,2
Closed circuit
12mA @12 VDC
Set switch S1-1 of Master
to NO.
CABLE
SHIELD
8
N/A
N/A
Keep cable length to a
minimum.
Table 7. J105B Outputs (Customer Interface)
J105B
Asserted
State
Circuit Ratings
Comments
Filament ON (K1)
3, 4
Closed
1 A @120 VAC
(10mA minimum)
K1 contacts close when filament
voltage is applied (STANDBY).
HV ON (K2)
5, 6
Closed
1 A @120 VAC
(10mA minimum)
K2 contacts close when high
voltage is applied to lamp.
QRO
7, 8
Closed
1 A @120 VAC
(10mA minimum)
K3 contacts close when QRO is
active.