beautypg.com

Specifications – Fluke RUSKA 2470 User Manual

Page 11

background image

Introduction

Specifications

1

1-3

Specifications

Low Range Piston

Pressure Range (Model 2465)

1.4 to 172 kPa (0.2 to 25 psi) psi gauge

Pressure Range (Model 2468)

1.4 to 345 kPa (0.2 to 50 psi) psi gauge

Pressure Uncertainty Rating

0.0010% (10 ppm) or 0.07 Pa (1.0E-05 psi), whichever is greater

(1,2)

Uncertainty Threshold

7 kPa (1 psi)

Resolution

1 ppm or 1 mg, whichever is greater

Precision (Typical Type A Unc.)

3 ppm (3)

Long Term Stability

3 ppm per year

Piston/Cylinder Material

440C Stainless Steel/Tungsten Carbide

Thermal Coefficient

1.5E-05 per deg. C

Sink Rate at Maximum Pressure

typical <2 mm per minute, maximum 4 mm per minute

Mid Range Piston

Pressure Range (Model 2465)

12 to 700 kPa (1.7 to 100 psi) absolute or gauge

Pressure Range (Model 2468)

12 to 1380 kPa (1.7 to 200 psi) absolute or gauge

Pressure Uncertainty Rating

0.0010% (10 ppm) or 0.07 Pa (1.0E-05 psi), whichever is greater

(1,2)

Uncertainty Threshold

35 kPa (5 psi)

Resolution

1 ppm or 1 mg, whichever is greater

Precision (Typical Type A Unc.):

3 ppm (3)

Long Term Stability

3 ppm per year

Piston/Cylinder Material

Tungsten Carbide/Tungsten Carbide

Thermal Coefficient

9.1E-06 per deg. C

Sink Rate at Maximum Pressure

typical <2 mm per minute, maximum 4 mm per minute

High Range Piston

Pressure Range

700 to 20680 kPa (100 to 3000) absolute or gauge

Pressure Uncertainty Rating

0.0030% (30 ppm) or 3.7 Pa (5.4E-04 psi), whichever is greater

(1,2)

Uncertainty Threshold

248 kPa (36 psi)

Resolution

1 ppm or 1 mg, whichever is greater

Precision (Typical Type A Unc.)

3 ppm (3)

Long Term Stability

3 ppm per year

Piston/Cylinder Material

Tungsten Carbide/Tungsten Carbide

Thermal Coefficient

9.1E-06 per deg. C

Sink Rate at Maximum Pressure

typical <2 mm per minute, maximum 4 mm per minute

Mass Set

Approximate Total Mass

17.8 kg

Approximate Carrier Mass

0.58 kg

Smallest Increment

5.9 gram

Mass Material

300 Series, Non-magnetic, Austenitic, Stainless Steel

(3)

Adjustment Method

Completely machined with no fill cavities

Mass Uncertainty

0.0005% (5 ppm) or 5E-07 kg, whichever is greater

Optional Fine Increment Trim Set

ASTM Class 1, 20g to 1 mg

(1)

Absolute mode uncertainty higher due to reference pressure sensor.

(2)

Approximate 95% level of confidence (Refer to Chapter 1, Accuracy, and to Calibration Report.)

(3)

Mass carrier composite construction 300 Series, Non-magnetic, Austenitic, Stainless Steel and other

non-magnetic material.