QED6 Users

Manual Supplement

11/10 1

Change #1,

56600

On page 1-10, replace the Output Power Measurement with
the following:

Output Power Measurement

Load Resistance ..................................................... 50

Ω ±1 % non-inductance

(< 10 µH), 160 W

Range

400 J ................................................................... 0-400.0 J

100 J ................................................................... 0-100.0 J

Resolution ............................................................... 0.1 J

Max. Vage

400 J ................................................................... 5500 V

100 J ................................................................... 1750 V

Max. Current

400 J ................................................................... 275 A

100 J ................................................................... 35 A

Measurement .......................................................... 400 J: 66 ±5 V

Trip Levels .............................................................. 100 J: 20 ±5 V

Pulse Width............................................................. 1-50 ms

Accuracy

400 J Range........................................................ ±2 % of reading

100-400 J ............................................................ ±2 Js

100 J Range........................................................ ±2 % of reading

±0.1 J

Change #2

On page 2-10, under the second step 4, replace the first bullet
with the following:
• Select the 400 joule high range (400 J) for defibrillator outputs

over 100 joules or for an unknown defibrillator output power.

On page 2-17, under step 5, replace the first bullet with the
following:
• Select the 400 joule high range for defibrillator outputs over 100

joules or for an unknown defibrillator output power.