Change #1, Change #2 – Fluke 2204520 User Manual
Page 2
QED6 Users
Manual Supplement
11/10 1
Change #1,
56600
On page 1-10, replace the Output Power Measurement with
the following:
Output Power Measurement
Load Resistance ..................................................... 50
Ω ±1 % non-inductance
(< 10 µH), 160 W
Range
400 J ................................................................... 0-400.0 J
100 J ................................................................... 0-100.0 J
Resolution ............................................................... 0.1 J
Max. Vage
400 J ................................................................... 5500 V
100 J ................................................................... 1750 V
Max. Current
400 J ................................................................... 275 A
100 J ................................................................... 35 A
Measurement .......................................................... 400 J: 66 ±5 V
Trip Levels .............................................................. 100 J: 20 ±5 V
Pulse Width............................................................. 1-50 ms
Accuracy
400 J Range........................................................ ±2 % of reading
100-400 J ............................................................ ±2 Js
100 J Range........................................................ ±2 % of reading
±0.1 J
Change #2
On page 2-10, under the second step 4, replace the first bullet
with the following:
• Select the 400 joule high range (400 J) for defibrillator outputs
over 100 joules or for an unknown defibrillator output power.
On page 2-17, under step 5, replace the first bullet with the
following:
• Select the 400 joule high range for defibrillator outputs over 100
joules or for an unknown defibrillator output power.