Power Source:

Four 1.5V DC (AA) batteries

Measurement
Method:

Oscillometric

Measurement Range:

Pressure: 40~280 millimeters mercury (mmHg)
Pulse: 40~199 beats/minute

Accuracy:

Pressure: ±3 mmHg
Pulse: ±5% of reading

Pressure Sensor:

Semi-conductor

Inflation:

Manual Inflation

Deflation:

Automatic Pressure release valve

Memory Capacity:

60 memories

Auto-shut-off:

1 minute after last button operation

Operation
Environment:

Temperature: 50°F~104°F (10°C ~40°C)
Humidity: 15% – 90% R.H.

Storage Environment:

Temperature: -4°F~158°F (-20°C~70°C)
Humidity: Less than 90% R.H.

Weight:

.43 lbs/193 grams (without batteries)

Arm Circumference:

Standard size adult arm cuff:
9”–13” (23 – 33 cm)

Dimensions:

5”(L) x 4.1”(W) x 2.7”(H)
126mm(L) x 105mm(W) x 68mm(H)

Internally Powered equipment
Type BF Equipment
IPx0-No special protection against the
ingress of water
Not suitable for use in presence of flammable
anesthetic mixture with air, oxygen or
nitrous oxide.
Continuous operation with short-time loading.

SPECIFICATIONS

Note: These specifications are subject to change without notice.

TROUBLESHOOTING

If any abnormality arises during use, please check the following points

Note: If the unit still does not work, contact HoMedics Consumer

Relations. Under no circumstance should you disassemble or
attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on
the warranty page.

SYMPTOMS

POSSIBLE

CAUSES

CORRECTION

Unit does not turn
on when the START/
STOP
button
is pushed.

Batteries have
run down.

Replace them with
two new AA alkaline
batteries.

Battery polarities
have been positioned
incorrectly.

Re-insert the batteries
in the correct
positions.

EE measurement
error symbol shown
on display or the
blood pressure
value is displayed
excessively low
(or high).

Cuff has been placed
incorrectly.

Wrap the cuff properly
so that it is positioned
correctly.

Did you talk or move
during measurement?

Measure again.
Keep arm steady
during measurement.

Shaking of the arm
with the cuff on.

E1 error symbol
shown on display

Air circuit abnormality.
Cuff tube may not be
plugged into monitor
correctly.

Check cuff connection.
Measure again.

E2 error symbol
shown on display

Inflation pressure
exceeding 300 mmHg.

Turn the unit off, then
measure again.

E3 error symbol
shown on display

Error determining
measurement data.

Measure again.