beautypg.com

Operating instructions, Disassemble – Precision Medical Oxygen Supply Manifold User Manual

Page 8

background image

7

Oxygen Supply

ManifOld

OPERATING INSTRUCTIONS

WARNING

Read this User manual before installing or operating the
Oxygen Supply manifold.
monitor patient and the oxygen flow on a frequent basis.
dO NOT direct flow of oxygen at any person, or

flammable material when cracking the oxygen cylinder.
Inspect Oxygen Supply manifold to ensure it is free of

oils, grease or other contaminations.


CAUTION

Inspect the Oxygen Supply manifold for visual damage

before use, DO NOT USE if damaged.

1. Attach one end of the Oxygen Supply Hose to the manifold

and the other end Oxygen Supply Hose to a 50 psi oxygen

source.

2. Attach up to six (6) oxygen supply masks to the Oxygen

Supply manifold tubing nipple / mask connections.

3. Place oxygen supply mask on patient/s.
. Turn the control knob until prescribed setting is aligned

with the Setting Indicating Arrow

on the Oxygen Supply

manifold.

dISASSEMBLE

1. Turn “OFF” the 50 psi oxygen source.
2. Remove the hoses from the oxygen source and the manifold.
3. Place Component back into Carrying Case.