7
Oxygen Supply
ManifOld
OPERATING INSTRUCTIONS
WARNING
Read this User manual before installing or operating the
Oxygen Supply manifold.
monitor patient and the oxygen flow on a frequent basis.
dO NOT direct flow of oxygen at any person, or
flammable material when cracking the oxygen cylinder.
Inspect Oxygen Supply manifold to ensure it is free of
oils, grease or other contaminations.
•
•
•
•
CAUTION
Inspect the Oxygen Supply manifold for visual damage
before use, DO NOT USE if damaged.
1. Attach one end of the Oxygen Supply Hose to the manifold
and the other end Oxygen Supply Hose to a 50 psi oxygen
source.
2. Attach up to six (6) oxygen supply masks to the Oxygen
Supply manifold tubing nipple / mask connections.
3. Place oxygen supply mask on patient/s.
. Turn the control knob until prescribed setting is aligned
with the Setting Indicating Arrow
on the Oxygen Supply
manifold.
dISASSEMBLE
1. Turn “OFF” the 50 psi oxygen source.
2. Remove the hoses from the oxygen source and the manifold.
3. Place Component back into Carrying Case.