SL Power Electronics MINT1065A Series User Manual
SL Power Electronics Equipment
SL Power Electronics, Corp..
6050 King Dr. Bldg. A
VENTURA, CA 93003 + 805-486-4565
MODEL NUMBERS: MINT1065BVWXYZ Series. Sample Model # is MINT1065B1275C01 which breaks
down as follows:
MINT
1
065
A
12
75
C
01
1
2
3
4
5 6 7 8
1 – Medical Internal Model # Prefix
2 – Signifies # of outputs: i.e. 1, 2, 3, 4, etc.
3 – Output Wattage: 065 is 65W
4 – Signifies Generational Differences such as Class I, Class II or Energy Star level changes. May be any Letter
A-Z. Where “A” represents Class I Power Supplies and “B” represents Class II Power Supplies.
5 – Output Voltage: Numeric indicator from 12 to 48. i.e. 12 = 12Vdc
6 – Output Connector options. 75 = 4 pin Tyco/AMP MTA156 or equivalent
7 – Input Connector options. C = 2 pin Tyco/Amp MTA156 or equivalent.
8 – Configuration: 01=Standard. 02-99 for modifications.
Unit is RoHS Compliant.
R A T I N G S
Input: 100-240 VAC, 1.3 A, 50-60 Hz (Korea input is 110V~, 1.27A)
Output: 12Vdc to 48Vdc at 5.25A to 1.35A.
MINT1065A BASE MODELS AND RATINGS
Model
Watts
Output
Model
Watts
Output
MINT1065A 12
63
12V/5.25A
MINT1065A 24
65
24V/2.7A
MINT1065A 15
65
15V/4.33A
MINT1065A 48
65
48V/1.35A
MINT1065A 18
63
18V/3.5A
Notes:
1. Maximum operating ambient is 40°C.
2. Maximum Relative Humidity 96%, no condensation.
CERTIFICATION: All models are certified to be in compliance with the applicable requirements of
UL/CSA/EN/IEC 60601-1.
CLASSIFICATION:
(5.1) Protection against electric shock = Class I
(In accordance with sub-
(5.2) Degree of protection against electric shock = Signal output or intermediate
clause 5 of IEC 60601-1)
(5.3) Protection against harmful ingress of water = Ordinary (no protection)
(5.5) Have not been evaluated for use in the presence of a flammable anesthetic mixture
with air, oxygen, or nitrous oxide. This evaluation is to be made on the end
equipment by the OEM.
(5.6) Mode of operation = Continuous
SAFETY DECLARATION: SL Power Electronics Corp (SLPE) declares under our sole
responsibility that all models listed above are in conformity with the applicable requirements of
EN60950 following the provisions of the Low Voltage Directive 2006/95/EC.
GROUNDING: The Protective Earth (ground) terminal must be bonded to Protective Earth in the host
equipment. Using the Protective Earth terminal on the supply for grounding the host equipment is not
recommended. A separate dedicated grounding point should be used.
Page 1 of 2
INSTALLATION INSTRUCTIONS
MINT1065A SINGLE OUTPUT SERIES