beautypg.com

SL Power Electronics MINT1065A Series User Manual

SL Power Electronics Equipment

background image




SL Power Electronics, Corp..
6050 King Dr. Bldg. A
VENTURA, CA 93003 + 805-486-4565

MODEL NUMBERS: MINT1065BVWXYZ Series. Sample Model # is MINT1065B1275C01 which breaks

down as follows:

MINT

1

065

A

12

75

C

01

1

2

3

4

5 6 7 8

1 – Medical Internal Model # Prefix
2 – Signifies # of outputs: i.e. 1, 2, 3, 4, etc.
3 – Output Wattage: 065 is 65W
4 – Signifies Generational Differences such as Class I, Class II or Energy Star level changes. May be any Letter

A-Z. Where “A” represents Class I Power Supplies and “B” represents Class II Power Supplies.

5 – Output Voltage: Numeric indicator from 12 to 48. i.e. 12 = 12Vdc
6 – Output Connector options. 75 = 4 pin Tyco/AMP MTA156 or equivalent
7 – Input Connector options. C = 2 pin Tyco/Amp MTA156 or equivalent.
8 – Configuration: 01=Standard. 02-99 for modifications.

Unit is RoHS Compliant.

R A T I N G S

Input: 100-240 VAC, 1.3 A, 50-60 Hz (Korea input is 110V~, 1.27A)
Output: 12Vdc to 48Vdc at 5.25A to 1.35A.

MINT1065A BASE MODELS AND RATINGS

Model

Watts

Output

Model

Watts

Output

MINT1065A 12

63

12V/5.25A

MINT1065A 24

65

24V/2.7A

MINT1065A 15

65

15V/4.33A

MINT1065A 48

65

48V/1.35A

MINT1065A 18

63

18V/3.5A

Notes:
1. Maximum operating ambient is 40°C.
2. Maximum Relative Humidity 96%, no condensation.

CERTIFICATION: All models are certified to be in compliance with the applicable requirements of

UL/CSA/EN/IEC 60601-1.

CLASSIFICATION:

(5.1) Protection against electric shock = Class I

(In accordance with sub-

(5.2) Degree of protection against electric shock = Signal output or intermediate

clause 5 of IEC 60601-1)

(5.3) Protection against harmful ingress of water = Ordinary (no protection)

(5.5) Have not been evaluated for use in the presence of a flammable anesthetic mixture

with air, oxygen, or nitrous oxide. This evaluation is to be made on the end
equipment by the OEM.

(5.6) Mode of operation = Continuous

SAFETY DECLARATION: SL Power Electronics Corp (SLPE) declares under our sole
responsibility that all models listed above are in conformity with the applicable requirements of
EN60950 following the provisions of the Low Voltage Directive 2006/95/EC.

GROUNDING: The Protective Earth (ground) terminal must be bonded to Protective Earth in the host
equipment. Using the Protective Earth terminal on the supply for grounding the host equipment is not
recommended. A separate dedicated grounding point should be used.

Page 1 of 2

INSTALLATION INSTRUCTIONS

MINT1065A SINGLE OUTPUT SERIES