Figure 3: Typical signal which can be seen at a detector preamplifier output, here from a

CZT500 detector. For optimum performance, the rise time should be in the
order of 100ns, and decay time constant of 50µs. This is drop to 36.8% of peak
value within 50µs or drop to half value within 34.6µs.