4. To return to normal weight display (gross weight),

remove any items from the weighing platform and

press the

TARE key again to re-zero the scale.

VII. Calibration
Calibration and adjustment are performed at the factory.

However, the weighing range can shift slightly during

shipping. Also, due to changes in local acceleration of gravity

around the world, adjustment is recommended when moving

the scale from one location to another. To check calibration:

1. Place the required calibration weight on the scale

and note the displayed value.

2. If the displayed value is not within acceptable

tolerance (see chart below), adjustment is required.

Model

Acceptable

Tolerance (±)

Calibration

Weight

Recommended

Accuracy Class

CARD-100

0.02g

100g

OIML M2

CARD-500

0.2g

500g

OIML M2

VIII. Adjustment

1. Press and hold the

MODE key until the display

shows “CAL”.

2. Press the

MODE key again. The display will flash

“CAL” followed by the calibration weight value (see

chart).

3. Place the calibration weight on the center of the

weighing platform. Make sure you are using the

correct weight for your specific model.

4. Once the weight is accepted, the display will show

“PASS” then return to normal operation.