beautypg.com

Labconco Purifie Horizontal Clean Benches 3970424 User Manual

Page 36

background image

Chapter 6: Maintaining Your Vertical Clean Bench

Product Service 1-800-522-7658

32

32

Calibration

1. Ensure the flow switch and alarm circuit board are installed and

operational.

2. Allow the enclosure to operate for at least two minutes.
3. If factory installed, the monitor will alarm at 30±10 fpm downflow velocity

with the average outflow velocity set at 85±10 fpm.

4. To change the factory setting, set the downflow velocity to the desired

alarm condition using the speed control adjustment procedure outlined in
Chapter 6. Once the alarm condition is set, use a small screwdriver to turn
the adjustment screw on the airflow switch counterclockwise (facing the
screw) until the “low” airflow red LED lights and the audible flow alarm
sounds.

5. Adjust the downflow velocity to the nominal operating point required. The

green “safe” LED should light.

6. Over time the HEPA filter will load and eventually slow the downflow

velocity. Once the alarm condition is met, simply increase the speed
control as outlined in Chapter 6 or replace the HEPA filter if the speed
control is maximized.

7. The table below lists typical alarm conditions based on normal operating

conditions. Typical alarm conditions are set at downflow velocities of 10
to 30 feet per minute below the normal operating conditions due to supply
air and exhaust air fluctuations, as well as room air cross drafts. Consult
your Safety Officer or Laboratory Manager for proper operating speeds.

Enclosure Operating

Downflow Speed

Alarm Condition Set Point

Speed

55 ± 10 fpm

25-35 fpm

45 ± 10 fpm

25-35 fpm