Care and Maintenance
Storage: After measurement, wrap cuff around gauge and bulb and store in zippered carrying case.
Cleaning: Digital Manometer: Your ADC
®
DIAGNOSTIX
TM
brand e-sphyg
TM
gauge requires minimal
care and maintenance.
The manometer may be cleaned with a slightly dampened soft cloth, but should not be dismantled
under any circumstances.
Cuff: Sponge with a damp cloth or alcohol pad. If necessary, it can be washed in cool water with
mild soap or detergent-disinfectant. Remember to remove bladder from cuff prior to washing. After
washing, it should be rinsed and allowed to air dry. NEVER IRON. Be sure to remove the bladder
and all components before cleaning the cuff.
Bladder and Bulb: Remove bladder from cuff and disconnect bulb from tubing and valves. Wipe
with a damp cloth. To help preserve it from cracking or discoloring, wipe with a soft cloth moistened
with ethanol.
Valves: Clean the surface with a dry fabric such as soft cotton material. Remove dust from filter
screen.
Carry Case: Wipe with a damp cloth and dry thoroughly with a dry cloth. If
necessary, it can be washed with cool water and mild soap. Do not allow to air dry.
Batteries: If the e-sphyg
TM
will not be used for an extended period of time, remove batteries.
Disinfection and Sterilization
Disinfecting: Glutaraldehyde-type liquid disinfectant may only be used on the cuff. Follow instruc-
tions for use provided with the glutaraldehyde product. Prolonged use of this disinfectant may
cause discoloration. DO NOT use glutaraldehyde-type liquid disinfectant on the digital gauge, bulb,
valves, or bladder.
Sterilizing: The blood pressure cuff may be gas sterilized. DO NOT use steam or heat to sterilize the
cuff, bulb, valve, or bladder. DO NOT attempt to sterilize the digital gauge.