beautypg.com

Operating the instrument, 1 remote control, 6 .3 operating the instrument – BUCHI Fraction Collector C-660 User Manual

Page 20

background image

6 Operation

20

C-660 Operation Manual, Version F

The mode line shows the navigation and editing mode you are in. It can be controlled with the selector
switch by rotating and pressing.

Parameters surrounded by lines cannot be
changed.

Fig. 6.3: Parameters surrounded by lines

Parameters highlighted with black background
can be selected and changed. The scroll bar on
the right indicates your position within the menu.

Fig. 6.4: Parameters highlighted with black background

With the selector switch you can navigate through the menu. You select the desired parameters by
turning the knob, and confirm your selection by pressing it. The cursor jumps into the editing mode or,
if it is already in the editing mode, into the next line (for program operation only in the stop mode). You
set the desired values by turning the selector switch. Press the selector switch in order to confirm the
new value. After 10 seconds without confirmation the cursor jumps back to the old parameters and
into the navigation mode.

6 .3

Operating the instrument

Turn on the fraction collector with the Power ON/OFF switch. The display shows the software version
for a few seconds. The main menu with the last used values and parameters now appears automati-
cally.
You can move through the menu by turning the selector switch. Select your parameters or compose a
new program.

6 .3 .1

Remote Control

If the fraction collector is part of a system controlled by SepacoreControl, “Remote Control” is indi-
cated on the display. In this mode no menu is accessible.
For this purpose set the address bus switch to position 1 or 2 (see Fig. 5.4). Now all settings have to
be done within the SepacoreControl software (see Installation Guide SepacoreControl).