Brookfield Bath, Model TC-602 User Manual
Page 30

29
cursor will automatically advance to the next field in the password entry display. Once the final
character is entered, rotate the knob and the following Calibration screen(s) will be available.
On Programmable Controllers, the Calibration User – External Probe Calibration screen will also
appear if an external probe is connected to the Controller.
To perform a calibration, go to the appropriate Calibration User screen and press the Select/Set Knob.
Rotate the Select/Set Knob until the offset value equals the difference between the probe temperature
reading and the reference temperature probe. Press the Select/Set Knob to accept the offset value.
NOTE: Allow the temperature reading at the probe to stabilize before making adjustments.
NOTE: The Calibration User display(s) remains available only while Controller power is On. If the
Controller in placed in Standby or main power is disrupted, the display(s) will have to be re-enabled
by entering the Calibration access password.