Appendix – Iris Sample Processing StatSpin® CritSpin Microhematocrit System User Manual
Page 15

11
Appendix
Appendix A - Specifications
Product No.
CS12 (supplied with RH12 Rotor)
CS22 (supplied with RH12 Rotor and CSD1 Digital Reader)
Model No.
M961
Speed
16,000 rpm + 250
Force
13,700 xg
Cycle
120 s: 120 seconds
Acceleration Time
Approximately 6 seconds
Deceleration Time
Approximately 10 seconds
Electrical
24 Volts, DC, 1.7 amp. Includes switching power supply for 100-240
VAC, 50/60 Hz
Dimensions
Height 6”/ 15.2 cm
Width 6.8”/ 17.3 cm
Depth 8.6”/ 21.8 cm
Weight 5 lbs/ 2.3 kg
Environmental
Indoor use
Altitude up to 2000m
Temperature 5ºC to 40ºC
Maximum relative humidity 80% for temperatures up to 31ºC decreasing
linearly to 50% relative humidity at 40ºC
Main supply voltage fluctuations not to exceed +/- 10% of the nominal
voltage
Transient over-voltages according to installation category II
Pollution degree 2
Product No.
CSD1
Refer to the CSD1 Operator’s Manual for specifications