2 inserting the reference electrode, Inserting the reference electrode – Metrohm Thin-Layer cell User Manual
Page 11
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
IC Equipment Thin-Layer cell
■■■■■■■■
5
3.2
Inserting the reference electrode
The measuring cell is supplied without electrodes. Reference electrodes
must be ordered separately.
Insert the reference electrode as follows:
Inserting the reference electrode
The reference electrode itself and one sealing ring are contained in the
packaging for the reference electrode.
1 Slide the sealing ring over the reference electrode.
2 Unscrew the fastening screw on the RE connector and remove it.
3 Insert the reference electrode into the opening with the flat side fac-
ing downward.
4 Retighten the fastening screw.
Special case: Ag/AgCl reference electrode
The Ag/AgCl reference electrode (6.1257.720) is, in contrast to the other
reference electrodes, permanently connected to the reference electrode
cable.