beautypg.com

2 inserting the reference electrode, Inserting the reference electrode – Metrohm Thin-Layer cell User Manual

Page 11

background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

IC Equipment Thin-Layer cell

■■■■■■■■

5

3.2

Inserting the reference electrode

The measuring cell is supplied without electrodes. Reference electrodes
must be ordered separately.

Insert the reference electrode as follows:

Inserting the reference electrode

The reference electrode itself and one sealing ring are contained in the
packaging for the reference electrode.

1 Slide the sealing ring over the reference electrode.

2 Unscrew the fastening screw on the RE connector and remove it.

3 Insert the reference electrode into the opening with the flat side fac-

ing downward.

4 Retighten the fastening screw.

Special case: Ag/AgCl reference electrode
The Ag/AgCl reference electrode (6.1257.720) is, in contrast to the other
reference electrodes, permanently connected to the reference electrode
cable.