KD Scientific Legato 100 Series User Manual
Page 68
Upgrading Legato 100 Series Software
1. Upload the latest software version to your desktop (format is filename.srec).
2. Disconnect all I/O devices and then connect the pump to PC using a USB cable.
3. On the Set up screen of the pump, press the Upgrade Software button twice to enter “Boot Loader
Mode”.
4. From the CD provided with the pump, open the Legato 100 Series Pump Updater Application.
5. Click Open Update File and browse to the file previously saved on your desktop and click Open.
6. Click Start Update.
7. After the update is complete, press Reboot to power cycle the pump.
68
Appendices
K
D
S
L
e
g
a
to
1
0
0
S
e
ri
e
s
U
s
e
r’s
M
a
n
u
a
l
KD Scientific Legato 100 Series Syringe Pump User’s Manual
Publication 5617-006 Rev D
✐
NOTE: Prior to upgrading, users will need
to install the KDS Legato 100 Series USB
driver as well as the bootloader driver file
(KDS Legato 100 Series-Bootloader.inf).
Reference “USB Virtual Comport Driver
Installation”.
phone
508.429.6809
fax
508.893.0160
web
www.kdscientific.com
NOTE: To exit boot loader mode without
upgrading the software, power cycle the
pump.
✐