Kds model 310 plus series, Specifications – KD Scientific 310 User Manual
Page 4
2
KDS Model 310 Plus Series
Specifications
Model
Model 310 Plus
Syringe Size
1–250 microliter
Electrical Rating
US model 115V~, 0.1A
CE model 230V~, 0.06A
Fuse
5x20 mm, 250V~ slow blow, 0.1A
Voltage Operating Range
US Model 100-120V~, 50/60Hz
CE Model 200-240V~,50/60Hz
Drive Mechanism
Microprocessor controlled stepper motor
(
1
⁄
2
microstep), leadscrew
Force
2 lbs (minimum)
Pusher Advance Per Half Step
1.58 microns
Minimum Stepping Rate
One
1
⁄
2
step every 4 seconds
Maximum Stepping Rate
916
1
⁄
2
steps/second
Flowrate Range
0.001–363.7 µl/min [0.5–250 µl syringes]
Dimensions
9x6x3 inch (23x15.25x8 cm) controller
7x1.7x2 inch (17.8x4.4x5.1 cm) injector
Weight
4.5 lb (2 Kg)
Atmospheric Specifications
Temperature
4˚C to 40˚C (40˚F to 104˚F)
Humidity
20% to 80% RH, non-condensing
Mode of Operation
Continuous
Classification
Class I
Pollution Degree
2
Installation Category
II
Output
N/A
Physiological Effects
N/A
Cooling Conditions
No special considerations
Mechanical Stability
No special considerations
Protective Packaging
No special considerations
Earth Terminals
No External connections required
Removable Protective Means
N/A
Supplier Name
KD Scientific Inc.
Address
84 October Hill Road, Holliston, MA 01746