IKA ULTRA-TURRAX Tube Drive User Manual
Page 9
9
032008
Page
CE-Declaration of conformity
9
Safety instructions
9
Unpack
10
Correct use
10
Useful facts
11
Commissioning
11
Error message
12
Accessories
12
Maintenance
13
Warranty
14
Technical data
14
We declare under our sole responsibility that this product corre-
sponds to the regulations 73/23/EEC, 89/336/EEC and 98/37/EEC
and conforms with the standards or standardized documents DIN
EN IEC 61010-1; DIN EN IEC 61326-1; DIN EN ISO 12100-1, -2 and
EN 60204-1.
For your protection
• Read the operating instructions in full before starting up and
follow the safety instructions.
• Follow the safety instructions, guidelines, occupational health
and safety and accident prevention regulations.
• Keep the operating instructions in a place where they can be
accessed by everyone.
• Ensure that only trained staff work with the appliance.
• The device must be supervised at all times when in operation.
• The device must only be operated with a tube attached; with no
tube there is a risk of injury from the rotating locating pin.
• Hearing protection must be worn when working with BMT-20 tubes.
• Wear your personal protective equipment in accordance with the
hazard category of the medium to be processed. Otherwise
there is a risk of:
- splashing liquids
- projectile parts
- body parts, hair, clothing and jewellery getting caught.
• Set up the appliance in a spacious area on an even, stable, clean,
non-slip, dry and fireproof surface.
• The feet of the appliance must be clean and undamaged.
• The appliance is not suitable for manual operation.
• The appliance may heat up when in use.
• Check the appliance and accessories beforehand for damage each
time you use them. Do not use damaged components.
• Ensure that the cover is screwed on tightly.
• Ensure that the tube is firmly attached to the bayonet lock
connector on the drive unit prior to operating the device.
• The tube must only be attached and removed whilst the motor is sta-
tionary.
• The IKA tubes must always be closed when the device is in ope-
ration. Switch off the device immediately if any material leaks
from the tube. Clean the equipment.
• Always open the tube carefully after use as the media in the tube
may heat up due to transfer of energy during operation, leading
to pressurization of the container: risk of material spraying, pro-
tective equipment must be worn.
• The temperature of the material must not exceed 40 °C.
• Only use tubes approved by IKA.
• Only process media that will not react dangerously to the extra
energy produced through processing. This also applies to any
extra energy produced in other ways, e.g. through light irradiation.
• Beware of the risk of flammable materials
• Do not operate the appliance in explosive atmospheres, with
hazardous substances or under water.
• The appliance starts not up again automatically following a cut in
Contents
Safety instructions
CE-Declaration of conformity
EN