Bayer HealthCare Rapidlab 800 User Manual

Page 110

background image

2Ć12

Rapidlab 800 Operator's Manual

Figure 2Ć6.ā Sample Type Menu

Select Micro Sample.

3.

Select

Micro Sample

and press

Enter

.

NOTE:ĄĂ

If the sample device is a syringe, the system tries to move the sample

automatically for routine analysis. If the syringe contains sufficient volume,
the system analyzes the sample automatically.

4.

Press

Analyze

.

The Position Sample Manually message box appears.

5.

Turn the sample pump clockwise to move the leading edge of the sample to
point A, as shown in Figure 2-7.

Figure 2Ć7.ā Sample at Point A on an 850

Point A

Sample Pump

6.

Press

Sample in Place

.

7.

When prompted, remove the sample device.