Accuracy:

The accuracy of the Ascensia

C

ONTOUR

Blood

Glucose Monitoring System has been assessed in a corre-
lation study conducted by healthcare professionals. In this
study, 108 fresh capillary blood specimens were tested
with the Ascensia

C

ONTOUR

Blood Glucose Monitoring

System by 108 untrained people with diabetes using three
test strip lots. The reference measurement was done on
the YSI Glucose Analyzer. Ascensia

C

ONTOUR

data compare

well with the results from the YSI method.

Regression Statistics (Plasma/Serum Reference):

plasma reference glucose range: 75 to 397 mg/dL

Precision:

A study was conducted with the Ascensia

C

ONTOUR

Blood Glucose Monitoring System with fi ve heparinized
whole blood specimens with glucose levels from 40 to
325 mg/dL. Multiple replicates (n=100) were tested using
multiple Ascensia

C

ONTOUR

Blood Glucose Meters and

one lot of Ascensia

C

ONTOUR

Blood Glucose Test Strips. The

following precision results were obtained:

Plasma/Serum:

Technical Information:

Technical Information

Lot

y =

Correlation

Coeffi cient (r)

1

1.02x – 0.12 mg/dL

0.972

2

1.03x – 4.34 mg/dL

0.963

3

0.95x – 0.41 mg/dL

0.974

Mean

SD

Coeffi cient

of variation

38 mg/dL

1.9 mg/dL

4.8%

101 mg/dL

5.1 mg/dL

5.1%

118 mg/dL

4.4 mg/dL

3.7%

205 mg/dL

6.8 mg/dL

3.3%

326 mg/dL

11.7 mg/dL

3.6%