2. Disabling Temperature Compensation

a.

As in Procedure to Select a Solution, pg. 10, select “USER”
mode.

b.

With “USER” selected, press

. If the display does not

show .00%/°C, hold

long enough to bring the tempco to

.00%/°C (see Figure 3).

c.

Press twice (3 times

if in TDS mode). Temperature
compensation is now disabled
(=0) for measurements in
USER mode.

3. User Programmable Conductivity to TDS Ratio

This feature allows you to select a custom conductivity to TDS conversion
ratio for USER mode measurements.

For example:

The conversion ratio range is 0.20-7.99 (ie., if conductivity

is 100 µS and TDS is 75 ppm, you would adjust to 0.75) (ref. Conductivity
Conversion to TDS, pg. 26).

a.

While in “USER” mode, press .

b.

Press twice (to skip over tempco adjustment), and

RATIO

” will appear (see Figure 4).

c.

Adjust with

or

keys until new conversion
ratio is displayed.

d.

Press

twice (to skip

over calibration adjustment)
to accept new conversion ratio.

You are now ready to measure samples with the new

conductivity/TDS ratio.

12

CAL

MCLR

MR

CAL

MCLR

T D S

CAL

MCLR

M S

MR

CAL

MCLR

Figure 3

COND

°C

% /

User

Figure 4

TDS

RATIO

User

VI. CALIBRATION

A. Calibration Intervals

Generally, calibration is recommended about once per month with
Conductivity or TDS solutions (ref. Calibration Intervals, pg. 19).

B. Rules for Calibration in the Ultrameter

1. Calibration Steps

a. Starting Calibration

Calibration is begun by pressing while measuring Conductivity or

TDS. Measuring continues, but the CAL icon is on, indicating calibration
is now changeable.

The reading is changed with the and

to match the known

value. The calibration for each of the 4 solution types may be performed
from either conductivity or TDS mode.

b. Calibration Steps

Depending on what is being calibrated, there may be 1, 2 or 3 steps to
the calibration procedures.

The

becomes an “ACCEPT” key. At each point, pressing

accepts the new calibration value and steps you to the next adjustment
(or out of CAL mode if there are no more steps).

To bypass a calibration step, just press

to accept the present value

as is.

13

In the first five sections, you have learned all
you need to make accurate measurements. The
following sections contain calibration, advanced
operations and technical information.

KCl, NaCl or 442

User

Cond

Gain only

Tempco, then Gain

Res

Done in conductivity

Done in conductivity or TDS

TDS

Gain only

Tempco, Ratio, then Gain

CAL

MCLR

M S

MR

CAL

MCLR

CAL

MCLR

CAL

MCLR