Appendix 1: Serial commands

M210300en

88

XPRES

Setting the pressure for mixing ratio and wet bulb

temperature calculations temporarily

XPRES pppp.pp

pppp.pp =

pressure (hPa)

The function and format of XPRES are the same as those of the PRES com-
mand except that by using XPRES the setting is valid only until a reset is
given or power is turned off or pressure is set to zero using XPRES. After this
the pressure stored using command PRES is valid again.

CDATE

Entering calibration date

CDATE xxxxxx

xxxxxx =

calibration date (000101...991231)

When the latest calibration date has to kept in memory, it is entered as fol-
lows:

>CDATE 940506

If the command is given without the date, the transmitter outputs the latest
calibration already in memory.

>CDATE
940420

The date can be given in any format; however, the maximum number of digits
is six.

DATE

Setting the date

DATE

For example, to enter a new date:

>DATE
Current date is 1993-01-30
Enter new date (yyyy-mm-dd) : 1993-06-12

When the current date is asked, the new date is passed with .

TIME

Setting the time