beautypg.com

Antenna-to-antenna isolation measurement 4-20, Antenna-to-antenna isolation measurement -20, Antenna-to-antenna isolation measurement – Anritsu S251C User Manual

Page 67

background image

Antenna-to-Antenna Isolation Measurement

Antenna-to-Antenna isolation measurement is accomplished in Insertion Loss Mode. Inser-
tion Loss Mode (+ 6 dBm) is used to drive the signal level large enough to be detected by
the second antenna system. The Insertion Gain Mode (–30 dBm level) does not output
enough power to provide an accurate measurement of Antenna-to-Antenna isolation .

Required Equipment

q

Site Master Model S251C

q

Test Port Extension Cable, Anritsu 15NNF50-1.5C (2 each)

q

Adapter, Anritsu 36NN50A

Procedure

Step 1. Press the

MODE

key.

Step 2. Select

INSERTION LOSS (+6 dBm) using the Up/Down arrow key and press

ENTER

.

Step 3. Set the frequency range for the antenna system. The table below indicates fre-

quency settings for common systems.

Step 4. Perform a full two port calibration of the Site Master (see page 3-4). The cali-

bration will provide a 0 dB reference level as shown in Figure 4-16.

4-20

Chapter 4 Measurements

Operation

Mode

System

F1 - Start Frequency

(MHz)

F2 - Stop Frequency

(MHz)

Insertion Loss

(+6 dBm)

CELLULAR

806

960

PCS

1850

1990

GSM

1800

1900

Figure 4-16.

Calibration Reference Level