beautypg.com

LumaSense Technologies ISR 6 Advanced User Manual

Page 41

background image

ISR 6 Advanced Manual

Data format UPP

Ò

· 41

Description

Command Parameters

Read

parameters

AApa

Answer: 15 decimal digits

DD . . . . . . . . . . . . . : Emissivity

(see em)

. . D . . . . . . . . . . . . : Response time (see ez)
. . . D . . . . . . . . . . . : Clear peak memory (see lz)
. . . . D . . . . . . . . . . : Analog output (see as)
. . . . . DD . . . . . . . . : Internal temperature (see gt)

. . . . . . . DD . . . . . . : Device address (see ga)
. . . . . . . . . 4 . . . . . . : Baud rate

(see br)

. . . . . . . . . . 0 , , , , : always 0
. . . . . . . . . . . DDDD : Ratio correction (see aw)

Serial number

AAsn

Output: XXXXX (hex 5-digit)

Read signal strength* AAtr

Answer: DDDD 4 decimal digit 0000 to 1500

Device type/

software version

AAve

Answer: VVMMJJ VV=54

MM=Month JJ=Year of software version

Communication
Module/ software
version in detail

AAvc

tt.mm.jj XX.YY
tt = day; mm = month; yy = year;
XX.YY = software version

Software version in
detail

AAvs

tt.mm.yy XX.YY
tt = day; mm = month; yy = year;

XX.YY = software version

*Read Signal Strength stands for product of emissivity, surface coverage and transmission of the measuring distance.

Note: the letter “l” means the lower case letter of “L”.

Additional instruction for the RS485 interface:

Requirements to the master system during half-duplex operation:

1. After an inquiry, the bus should be switched into a transmission time of 3 bits (some

older interfaces are not fast enough for this).

2. The pyrometer's response will follow after 5 ms, at the latest.
3. If there is no response, there is a parity or syntax error and the inquiry has to be

repeated.

4. After receiving the response, the master has to wait at least 1.5 ms before a new

command can be entered.