beautypg.com

SKC 228-9530 QuickTake 30 User Manual

Page 11

background image

10

• To navigate presets and displays:
Press

 or 

to scroll through presets or displays.

• To select a run time preset and run the pump:
Scroll to desired run time preset.
Select preset by pressing [



]. The pump will start to run.

• To run the pump:
Press

[



].

• To place the pump in hold:
Press

[



].

• To repeat a sample run:
From Done, reset (rSEt), or sampling error (SErr), press [



] to return to run

time preset. Press [



] to sample.

To turn off LED:
The pump automatically turns off LED aft er it is idle for 4 minutes. User can

reactivate the LED by pressing and holding

.

• To interrupt a run, terminate a sample, or reset a pump:
From Hold, press

. Sample reset (rSEt) will display. Press [



] to reset the

pump or

to ignore the reset and return to Hold.

• To set fl ow or calibrate:
See Calibration and Sett ing Flow for details.
1. Connect a calibrator to the pump inlet.
2. Scroll to any run time preset. Press and hold CAL for 2 seconds to enter

calibration mode (pump will start running).

3. Press

 or  to change fl ow rate until desired fl ow appears on the

calibrator (fl ow rate does not display on LED).

4. Press

and Stor will display. Press [



] to save sett ing or

to ignore

changes.

Selecting the

Available Presets

Press  or 

Press 

5

4

3

2

1

CONT

INTT

END

999

998

997

Press 