Section 4: Installing Accessories

04/11

26

51255 Rev. 2

Note: This truck scale has been designed to provide protection from the effects

of moisture. The load cells have been calibrated with the cable attached, and

therefore the cable should NOT be cut.
The cable is connected directly to the balance box or SSC through a sealed bushing

which MUST be tightened properly to keep water/moisture out of the box.
All cabling should have a drip loop at the cell or box entry location to help prevent

water entry.
On all boxes, particularly stainless steel, the black plastic fittings have “O” rings

that can be forced out of position if the bushing itself is not tight.
To prevent this, first tighten the inner nut securing the bushing in the hole, then

insert cable and carefully tighten the gland until it is very snug.
Do not over-tighten where bushing turns.
Secure the cover.

Note: Balance boxes must have one (1) ground rod in the pit for proper grounding.

C A UTION

Inadequate grounding will prevent the surge

voltage protection from functioning properly.

1. Instrument cable connection, balance box

The two (2) cables from the two (2) center section boxes will enter the 14478

instrument SVP and terminate there. The cable from the instrument will

connect at 14478 instrument SVP as well. Prepare the cable ends in the

standard manner. Use Appendix II for wiring instructions of all pit balance

boxes. Connect the instrument interface cable to the instrument in the scale

house per the instructions in the appropriate instrument service manual.

2. Adjusting cells/sections

Try to install load cells of matching outputs in sections to reduce side-to-side

errors. When calibrating, place weights directly over the cell or directly on the

section being tested. Adjust the potentiometers for the correct cell or section

to compensate for differences.