OPERATION
Constant Current:
0 to selected full scale current
Prog. Accuracy: .1% of setpoint ±5ma
Resolution(IEEE): .0015% of selected full scale
Constant Resistance:
Constant Resistance mode operates in
Amps/Volt, IEEE
units entered in ohms or A/V
Prog. Accuracy: 1% of setpoint +- 5 ma
Resolution(IEEE): .0015% of selected full scale
Constant Voltage:
0 to selected selected full scale
Prog. Accuracy: 1% of setpoint +- 250mv
Resolution(IEEE): .0015% of selected full scale
Constant Power:
0 to full scale power
Prog. Accuracy: 1% of setpoint +-250mw
Resolution(IEEE): .0015% of full scale power
ANALOG MODE
Ext. Prog:
0 to 10 Volts input yields 0 to selected full scale Input
Impedance: 330k Ohms
PULSE MODE
Frequency: 1Hz to 5kHz
Duty Cycle: 0 - 100%
Minimum Pulse Width: 125usec
Adjustable Slew Rate:
Max: 0 to full scale in 10µS - adjustable
OUTPUT SIGNALS
Current Sample Output:
Scaling: 10 Volts = selected full scale
Accuracy: ±0.5% of selected full scale
MISCELLANEOUS
AC Input:
User Selectable 100VAC,120VAC, 200VAC, 240VAC,
±10%, 48 - 62 Hz @ 350W
Ambient Temp:
0°C to 40°C
PROTECTION
Current Limit:
Analog Models: Approximately 105% of selected full scale current
Range(IEEE): 0-105% of selected full scale
Resolution(IEEE): 0.4% of selected full scale
Voltage Limit:
Analog Models: Load disconnect at 105% of selected full scale
voltage
Range(IEEE): 0-105% of selected full scale
Resolution(IEEE): 0.4% of selected full scale
Power Limit:
Range(IEEE): 0 - 6300 Watts
Resolution(IEEE): 0.4%
Thermal:
Load disconnect at internal temperature of 105°C
Undervoltage:
Load inhibited at less than1 Volt, when enabled
COMMUNICATION CHANNEL READBACK
Current:
Resolution: .0015% of Selected Full Scale
Accuracy(Range): 0.1% ±5ma
Voltage:
Resolution: .0015% of Selected Full Scale
Accuracy(Range): 0.1% ±5ma
Power:
Resolution: 1 Watt
Accuracy: 0.1% ±5ma
Communication Modes:
IEEE 488
RS 232
Ethernet (HTTP, TCP, & Telnet)
I
NNOVATIVE
T
EST
S
YSTEMS
... G
LOBAL
T
EST
S
OLUTIONS
XBL S
ERIES
6000
G
ENERAL
S
PECIFICATIONS
(
PRELIMINARY
)
•
Excellent Programming Accuracy
•
High Speed Programmable Slew Rate
•
Ethernet Control
•
Closed Box Calibration
•
RBL Command Support
•
Pulse Load Shaping
•
Full Range Switching
•
19 inch Rack Mount
The XBL Dynaload Series features 800, 2000,4000, and 6000 watt models with wide range voltage inputs and sophisticated
computer programming via GPIB, Ethernet, or RS232. These units feature an integrated web page for local system operation
and control.
Individual models are available for specialized low voltage high current applications. High voltage models up to 1000 volts are
also available.
All models include easy to apply master slave parallel capabilities and closed box calibration. Higher power models
incorporate variable speed forced air cooling to assure a quiet environment.