Performing correlation in ooilibsplus, Accessing the correlation window – Ocean Optics LIBS2500plus User Manual
Page 40

3: Operation
30
166-00000-000-02-0110
2. Select a line from the element list, then click the Add button. Repeat this step until you have
loaded all the desired lines.
3. Modify any properties, the L Wave, the R Wave, or set the Wavelength value for custom waves.
4. Click the Update button to save the changes.
5. Click the Scan button repeatedly.
You should begin to see data appear in the Lines to Analyze section.
Performing Correlation in OOILIBSplus
Follow the instructions in the sections below to perform a correlation in OOILIBSplus:
Accessing the Correlation Window
Adding Samples to the Correlation Library
Saving the Correlation Library
Accessing the Correlation Window
►
Procedure
1. Start the OOILIBSplus software.
2. Obtain a few samples to make sure that the system is working correctly.
- Apex 785 Raman (1 page)
- STS-UV (2 pages)
- TR2 Engineering Note (4 pages)
- SAD500 Communications and Control (19 pages)
- Red Tide USB650 Install (26 pages)
- Fiber Optic Termination Kit (6 pages)
- Transmissive pH Probe (10 pages)
- Remora (42 pages)
- PlasCalc (59 pages)
- Correcting Device Driver Issues (8 pages)
- ecoVis Krypton Light Source (16 pages)
- LPC-500CM (28 pages)
- HPX-2000 (24 pages)
- ADC1000-USB (27 pages)
- Torus Operating Instructions (30 pages)
- ADC2000-PCI (13 pages)
- Sensors for Real-Time Analysis (2 pages)
- IDRaman reader (2 pages)
- DH-2000-CAL (30 pages)
- QE65 Pro (32 pages)
- Collimating Lenses (2 pages)
- D1000 (2 pages)
- IDRaman mini (2 pages)
- HR2000CG-UV-NIR (42 pages)
- Cool Red (1 page)
- HL-2000 (20 pages)
- XE-1 Xenon (6 pages)
- USB-ADP Serial Adapters (3 pages)
- CHEM2000 (36 pages)
- Deuterium-Halogen Calibration Light Source (19 pages)
- NeoFox Engineering Note (30 pages)
- QE Pro (2 pages)
- OOIBase32 (140 pages)
- AR-1 Argon (6 pages)
- pH Sensor Patches, Probes and Cuvettes (36 pages)
- SpecLine Offline Spectroscopy (60 pages)
- HL-2000-HP-232 (26 pages)
- External Triggering Options Instructions for Spectrometers with Firmware Version 3.0 and Above (16 pages)
- Breakout Box (10 pages)
- USB-ISS-UV_VIS (4 pages)
- ISS-UV_VIS (6 pages)
- OOIColor (14 pages)
- LS-1 Series (12 pages)
- Apex Install (24 pages)
- DH-2000 (34 pages)