General specifications
Model 61-362
Characteristics
Description
Display
3 ½ Digit LCD display
Display Count
3200 count, maximum reading 3199
32 segment analog bar graph
Over range Indication
“OL” displayed
Sampling Rate
2 time/second
Operating Environment:
Relative Humidity
0
°C to 50°C (32°F to 122°F) 70%RH
Temperature Coefficient
0.05X x (accuracy) per
°F (32°F to 65°F, 85°F to 122°F
Storage Environment:
-20
°C to 60°C (0°F to 140°F) at <70 relative humidity
Power source:
9V Battery (NEDA 1604)
Battery Live:
200 hours typical (alkaline)
Low Battery Indicator:
symbol indicates low battery voltage
Auto Power Off mode
Approximately 25 minutes
0.5A/250V fast acting fuse Type LA-3895
A protection Fuse
10A/600V fast acting fuse, Type LA-3897
Dimensions
5.75” H X 2.75” W X 1.5” D {without holster}
Weight:
Approximately 18.0 oz. including battery
Safety
UL1244, and Design to comply with IEC 1010-1 Cat III
RANGES and ACCURACY SPECIFICATION
61-362
Function Setting
Ranges
Accuracy
3.200V/32.00V/320.0V/ 600V
AC Voltage
50Hz to 500Hz
2.0% ± 4 digits
320.0mV/32.00V/320.0V/600V
1.2% ± 1digit
DC Voltage
3.200V
0.8% ± 1digit
320µA/3200 µA/32.00mA/320mA
2.5% ± 4 digits
10A,
AC Current
50Hz to 500Hz
3.5% ± 4 digits
320µA/3.20mA/32.00mA 2.0%
± 1 digit
DC Current
10A,
3.0% ± 1 digit
320.0
Ω
2.0% ± 3 digits
3.200k/32.00k/320.0k
1.5% ± 3 digits
3.200M
Ω
2.5% ± 3 digits
Resistance
32.00M
Ω
5.0% ± 5 digits
Continuity
beep on < 100
Ω
Not Specified
Diode Check
2V DC max
10%
± 2 digit
AC Converter:
Average responding, RMS Calibrated to Sine Wave
Overload Protection:
AC and DC Volts: 600V DC or AC rms .
Resistance, Diode, Continuity: 500V DC or AC rms
mA input: .5A/250V DC/AC rms
Amps input: 10A/600V DC/AC rms
Form number 61362