beautypg.com

LifeSource UA-100 User Manual

Page 8

background image

LifeSource Health Line (Toll-Free): 1-888-726-9966

E-10

LifeSource Health Line (Toll-Free): 1-888-726-9966

How to Check your Home Blood Pressure Kit for
Accuracy

1

.

Detach manometer from bladder tube. Hold manometer in a
vertical position.

2.

If the needle on the dial rests within the zero indicator on the
face plate, your unit is guaranteed to be accurate.

3.

If the needle rests outside the zero indicator, it should be
recalibrated.

4.

To recalibrate, insert the small screwdriver into the hose
connector end of the manometer. Turn the screw driver in either
direction without going all the way around the dial to get the
needle back to the zero indicator.

5.

If the unit cannot be recalibrated using this method, please
contact A&D Medical at 1-888-726-9966.

Two-Year Calibration Warranty

A&D Medical warrants, to the first retail purchaser of the Manometer
(the Product), free accuracy adjustment for a period of two years if
the Product becomes inaccurate under normal conditions of use. See
enclosed warranty card for complete details.

CHECKING FOR ACCURACY

E-11

MAINTENANCE INSTRUCTIONS

Basic Care of Your Home Blood Pressure Kit

With proper care, your Home Blood Pressure Kit will provide many
years of reliable service. To insure a well functioning unit, follow
these basic rules:

• Handle the pressure manometer with care. Do not drop or jar.

The needle should indicate zero when cuff is fully deflated.

• Always deflate cuff before storing.

• Avoid hitting objects against surface of stethoscope.

• Store all items, including instruction booklet, in the convenient

storage case provided.

A&D Medical has checked your blood pressure unit to assure reliability
and accuracy prior to shipment. However, as with any sensitive instru-
ment subject to repeated use, we recommend your blood pressure unit
be checked periodically for proper calibration to ensure accurate blood
pressure readings.