Invacare IRC5P User Manual
Page 14

SECTION 4—TYPICAL PRODUCT PARAMETERS
14
*NOTE: Stated concentration levels achieved after initial warm-up
period (approximately 30 minutes).
Safety System:
Current overload or line surge shutdown.
High temperature compressor shutdown.
High Pressure Alarm w/compressor shut-
down. Low Pressure Alarm w/compressor
shutdown. Battery Free Power Loss Alarm.
SensO
2
Oxygen System (SensO
2
Model)
Low Flow Alarm.
Width:
15 in. ± 3/8 in (38.1 cm ± 1 cm)
Height:
23 in. ± 3/8 in (58.4 cm ± 1 cm)
Depth:
12 in. ± 3/8 in (30.5 cm ± 1 cm)
Weight:
Perfecto
2
5: 43 lbs ± 2 lbs (19.5 kg ± 1 kg)
Shipping Weight:
Perfecto
2
5: 48 lbs ± 2 lbs (21.8 kg ± 1 kg)
Operating Ambient
Temperature:
50°F - 95°F (10°C - 35°C) at 20-60% relative
humidity
Exhaust:
Perfecto
2
5: Less than Ambient
+ 35°F (+19°C).
Oxygen Output:
Perfecto
2
5: Less than Ambient + 6°F (+3°C)
Cabinet:
Impact Resistant flame-retardant plastic cab-
inet that conforms to UL 94-V0.
Regulatory Listing:
Double Insulated Product ETL certified
complying with UL 1097 ETL certified to
CSA C22.2 No. 68.
Electrical:
No extension cords.
Placement:
No closer than three inches from any wall,
furniture, draperies, or similar surfaces.
Tubing:
7 ft cannula with a maximum 50 ft of
Crush-Proof Tubing (DO NOT pinch).
Relative Humidity:
20 to 60%
Time of Operation:
Up to 24 hours per day.
Recommended Storage and
Shipping Temperature:
-20°F to 150°F (-29°C to 65°C), at 15-95%
relative humidity.