beautypg.com

Treatment instructions, Ꮨ danger – Invacare IRC 1730 User Manual

Page 19

background image

SECTION 2—OPERATING INSTRUCTIONS

Part No 1150755

19

Stratos™Portable Plus

10. To use with battery, charge for 6 hours. Refer to 

Conditioning the Battery on page 23.

11. To use the nebulizer system with the AC power cord, 

connect the power cord plug into the power outlet on the 
side of the nebulizer compressor system. Connect the 
power cord to the AC adapter and insert the plug from the 
AC adapter into the electrical outlet.

12. To use the nebulizer system with the DC auto adapter, 

connect the DC auto adapter to the power outlet on the 
side of the nebulizer compressor system. Connect the auto 
adapter plug of the DC cord to the auto adapter receptacle.

DANGER

To avoid the risk of possible electrical shock, the nebulizer

system MUST be operated on the specified power source.

Treatment Instructions

NOTE: For this procedure, refer to FIGURE 2.2 on page 20.

NOTE: The Invacare Stratos Portable Plus Nebulizer System has 
been designed for intermittent use. Normal treatment time should be 
five to ten minutes. Turn off the nebulizer system after 30 minutes 
of use and allow the device to cool down 30 minutes before turning 
it on again.