beautypg.com

Oxygen purity, Senso, Sensor feature – Invacare IRC5LXO2 User Manual

Page 16

background image

16

INITIAL STARTUP OF THE CONCENTRATOR

NOTE: Concentrator may be used during the initial start warm-up time

(approximately 30 min.) while waiting for the GREEN O

2

purity light to illuminate.

When the unit is initially turned on, the GREEN power LED (I/O) will

illuminate. Within 5 to 30 minutes after initial startup, a GREEN light will

illuminate in the Oxygen Purity section of the Control Panel (FIGURE

7) if concentration is within specification. Any other SensO

2

® indicator lights

illuminating will signal a condition that requires a call to the supplier.

OPERATING INSTRUCTIONS

EXPLANATION OF OXYGEN PURITY INDICATOR

LIGHTS (FIGURE 7):

GREEN LIGHT ( ) - Normal Operation
YELLOW LIGHT ( ) - Call Supplier IMMEDIATELY. You may

continue to use the concentrator unless instructed otherwise by your

supplier. Be certain that BACKUP OXYGEN is nearby.
RED LIGHT ( )- Total unit shutdown. Switch IMMEDIATELY

to BACKUP OXYGEN supply. Call Supplier IMMEDIATELY.
GREEN LIGHT - with YELLOW LIGHT FLASHING - Call

Supplier IMMEDIATELY. Oxygen sensor malfunctioning; you may

continue to use the concentrator.

O

2

FIGURE 7 - O

2

DISPLAY CONTROL PANEL

Call Supplier

Normal

Below Normal

Oxygen Purity

RED

YELLOW

GREEN

O

2

Auto Shut - Down

Use Backup

SENSO

2

® OXYGEN PURITY INDICATOR-

IF YOUR UNIT FEATURES THE O

2

SENSOR

FEATURE

This feature monitors the purity of oxygen generated by the oxygen

concentrator. If purity falls below factory preset standards, indicator

lights on the control panel will illuminate.

This manual is related to the following products: