SECTION 3—OPERATION

Part No. 1078984

13

Stand Up Patient Lift

Activating a Mechanical Emergency

Release

Primary Emergency Release

NOTE: For this procedure, refer to FIGURE 3.2.
NOTE: This procedure will bring the boom down if the hand control 
is not functioning properly.
To activate the primary emergency release, insert the tip of a 

pen into the Emergency Down hole () in the control box.

Secondary Emergency Release

NOTE: For this procedure, refer to FIGURE 3.3.
NOTE: All patient lift actuators are equipped with a mechanical 
emergency release. The mechanical release will enable the actuator to 
retract without power.
NOTE: Use the primary emergency release first before using the 
secondary emergency release procedure. This procedure should only 
be used if the primary emergency release procedure is not functioning 
or is unreachable.
NOTE: The lift MUST be under a load for the mechanical release to 

function.
To activate the secondary emergency release, pull up on the 

RED emergency grip and pull down on the boom at the same 

time.

Charging the Battery

NOTE: For this procedure, refer to FIGURE 3.4.
NOTE: Invacare recommends the battery be recharged daily to 
prolong battery life. 
NOTE: An audible alarm will sound when battery is low.
1.

Lift UP on the handle on the back of the battery.

2.

Lift the battery up and away from the control box.

CAUTION

Make sure there is an audible click when mount-

ing battery on the battery charger. Otherwise,

injury or damage may occur.

3.

Place the battery on the battery charger. Push the top of the 
battery against the mounting bracket until there is an 
audible click.

NOTE: The charge LED will illuminate. When charged, the LED 
will stop illuminating.
NOTE: It will take approximately four hours to charge a battery that 
requires a full charge.
4.

Lift UP on the handle on the back of the battery.

5.

Lift the battery up and away from the battery charger.

CAUTION

Make sure there is an audible click when mount-

ing battery on the battery charger. Otherwise,

injury or damage may occur.

6.

Place the battery on the control box. Push the top of the 
battery against the mounting bracket until there is an 
audible click.

FIGURE 3.2 Primary Emergency Release

FIGURE 3.3 Primary Emergency Release

FIGURE 3.4 Primary Emergency Release

Emergency Down

Hole

Control Box

RED Emergency Grip

Battery Charger

(STEP 3)

Control Box

(STEP 6)

An audible “click”

will be heard when

properly installed

(STEPS 3 and 6)

Handle

STEPS 1, 2, 4 and 5)

Battery