RJS Inspector D4000 Auto Optic (FIRMWARE version A.05 and Earlier) User Manual
Page 17

Operator
’
s Guide
Inspector 4000
™
002-5564
RJS, Minneapolis, MN
10
Verifying that Unit is Calibrated
Selecting CALIB sets up the mode to measure reflectance. This also allows you to
check the calibration of the unit by placing the wand on a template (white or black
field) and verifying that the reflectance value matches the stated values.
Verify that reflectance values are within 5% for white and 2% for black. If the values
are not within these tolerances you need to calibrate the unit; otherwise, press Enter
and Select at the same time to navigate back to the Main Menu.
Calibrating the Unit
After selecting Calib from the main menu, the reflectance menu will display.
Inspector 4000
Reflectance XX%
From the
reflectance
menu press Enter, the Calibrate menu
will
display:
Calibrate
When the “Calibrate” menu is displayed, proceed with the following calibration
steps
:
Place Optics on
White/Enter Key
Place the
wand
on the white section of the calibration plaque with the shoe
guide
flat
against the surface. Press Enter and hold auto-optic scanner head still until you hear a
“chirp” sound.