beautypg.com

Teds configuration: strain-gage load cell, Teds configuration: accelerometer – Measurement Computing IOtech 640 Series User Manual

Page 96

background image

C-4 TEDS

928693

Appendix

C

The Basic TEDS contains the manufacturer’s identification, model identification, version letter, and the serial
number, which are common to all types of sensors. The Standard TEDS, however, contains data unique to the
sensor’s principle of operation. For instance, a piezoelectric accelerometer differs from a strain gage load cell, so the
Standard TEDS EEPROM contains some common and some unique information. Both sensor TEDS list Calibration
Date, Measurement Range, and Electrical Output, but they differ in as many as 12 other parameters. The
accelerometer has 6 parameters that the strain gage lacks, and the strain gage has 6 parameters the accelerometer
lacks. The User Area contains information such as Sensor Location, Calibration Due Date, and Calibration Table.

TEDS Configuration: Accelerometer

Basic TEDS

Manufacturer ID

25

Model ID

5226

Version Letter

A

Serial Number

11521S

Standard TEDS Calibration Date

Jan. 6, 2004

(Subtemplate)

Sensitivity @ Reference

1.12E+03 mV/g

Reference Frequency

110.0 Hz

Reference Temperature

25˚C

Measurement Range

±50 g

Electrical Output

±5 V

Quality Factor
Temperature Coefficient

-0.27%/˚C

Direction (X, Y, Z)

X

User Area

Sensor Location

Matrix A: B-3

Calibration Due Date

Jun. 15, 2004

Fig. C.04. Table 2. Example for an Accelerometer.





TEDS Configuration: Strain-Gage Load Cell

Basic TEDS

Manufacturer ID

56

Model ID

350A

Version Letter

E

Serial Number

150-00878T

Standard TEDS

Calibration Date

Mar. 15, 2004

Measurement Range

±250 lb

Electrical Output

±5 mV/V

Bridge Impedance

500

Measurement Range

±75 g

Excitation, Nominal

24 Vdc

Excitation, Minimum

10 Vdc

Excitation, Maximum
Response Time

Calibration TEDS Calibration Table

-2.06 mV/V, -150 lb
-1.02 mV/V, -75 lb
0.03 mV/V, 0 lb
0.95 mV/V, 75 lb
1.89 mV/V, 150 lb

User Area

Sensor Location

L23-3

Calibration Due Date Jun. 15, 2004

Fig. C.04. Table 3. Example for a Strain-Gage Load Cell.