Appendix, P = 10, U = c + 0.6log – INFICON MPG40x ATM to Ultra-High Vacuum Gauge User Manual

Page 29: Measuring signal vs. pressure

background image

tina48e1-b (2008-10) MPG400-401.om

29

Appendix



p = 10

1.667U-d

U = c + 0.6log

10

p

p U c

d

[mbar]

[Torr]

[Pa]

[V]
[V]
[V]

6.8

6.875

5.6

11.33
11.46
9.333

where p pressure

U measuring

signal

c, d constant (pressure
unit

dependent

valid in the range

5×10

-9

mbar < p

3.8×10

-9

Torr < p

5×10

-7

Pa < p

<1000 mbar
<750 Torr
<1×10

5

pa


1E+04

1E+02

1E+00

1E–02

1E–04

1E–06

1E–08

0.0 0.5

1.5

2.5

3.5

4.5

5.5

6.5

7.5

8.5

9.5

10.5

sen

sor

er

ror

un

der

ra

nge

sen

sor

er

ror

ove

rr

ang

e

Pa

mbar

torr

Measuring signal U [V]

Pressure p


Meas. signal U [V] Pressure p [mbar]

Pressure p [Torr]

Pressure p [Pa]

<0.5 Sensor

error

0.5 … 1.82

Underrange

1.82

2.0
2.6
3.2
3.8
4.4
5.0
5.6
6.2
6.8
7.4
8.0
8.6

5.0Ч10

-9

1.0Ч10

-8

1.0Ч10

-7

1.0Ч10

-6

1.0Ч10

-5

1.0Ч10

-4

1.0Ч10

-3

1.0Ч10

-2

0.1
1.0

10

100

1000

3.8Ч10

-9

7.5Ч10

-9

7.5Ч10

-8

7.5Ч10

-7

7.5Ч10

-6

7.5Ч10

-5

7.5Ч10

-4

7.5Ч10

-3

7.5Ч10

-4

0.75

7.5

75

750

5.0Ч10

-7

1.0Ч10

-6

1.0Ч10

-5

1.0Ч10

-4

1.0Ч10

-3

1.0Ч10

-2

0.1
1.0

10

100

1000

1.0Ч10

4

1.0Ч10

5

8.6 … 9.5

Overrange

9.5 … 10.5

Sensor error (Pirani defective)

A:

Measuring Signal vs.

Pressure

Conversion formulae

Conversion curves

Conversion table